32
1036100102L02 1904V002
Usage
12 Maintenance
12.1 Maintenance schedule
Prior to working on the unit or in case of danger, disconnect it from the mains.
Unit without membrane-drying unit
NOTICE
Risk of damage to the unit due to blocked filters
Continuous running due to reduced delivery. Damage to the unit due to burst filters.
i
Replace filters in accordance with the maintenance schedule.
Maintenance
interval
Maintenance work
At regular inter-
vals
i
Drain the condensate from the pressure vessel by opening the condensate
drain valve (the interval can vary depending on the ambient conditions and
method of working; empty it daily with high air humidity).
Annually
i
Replace the air intake filter - every six months given a high concentration of
dust.
Every 4 years
i
Replace the vibration dampers.
In accordance
with national di-
rectives
i
Check the safety valve.
i
Carry out repeat safety tests (e.g. pressure vessel test, electrical safety test)
in accordance with the national directives.
Unit with membrane-drying unit
NOTICE
Risk of damage to the unit due to blocked filters
Continuous running due to reduced delivery. Damage to the unit due to burst filters.
i
Replace filters in accordance with the maintenance schedule.
Maintenance interval
Maintenance work
At regular intervals
i
Empty the collector tray under the membrane drying unit / condensate
drain (the interval may vary depending on the ambient conditions and
method of working; empty it daily if the humidity is high).
Annually
i
Replace the air intake filter - every six months given a high concentra-
tion of dust.
i
Change the 3 µm / 0.01 µm fine filter cartridge.
i
Replace the sintered filter.
Every 4 years
i
Replace the vibration dampers.
In accordance with na-
tional directives
i
Check the safety valve.
i
Carry out repeat safety tests (e.g. pressure vessel test, electrical safety
test) in accordance with the national directives.
EN
Summary of Contents for H2A-160M
Page 2: ......