background image

2  MEASUREMENT PROCEDURE 

2.1  Start Up 

2.1.1  Temperature Probe 

Load the temperature logger so

ft

ware and connect 

the temperature logger to the PC using the serial 
cable provided. Con

fi

gure the logger to take 

temperature readings at frequent intervals (these may 
be far more frequent that the RAD7 test cycles.)  

Choose the second (external) temperature sensor.  
Connect the temperature probe to the logger and 
note that when you hold the probe the indicated 
temperature rises. 

When everything is set, begin recording.  

Th

e logger’s 

LED may 

fl

ash periodically.  Once the logger has 

begun running, you may remove the serial cord from 
both the logger and the PC. 

Warning!

 

Make sure that previous temperature data 

has been downloaded before launching the logger. 

Th

e launching process may erase previous data. 

If not already done so, insert the probe into the RAD 
AQUA. 

2.1.2  Start Measurement 

Start the water 

fl

owing. Note that, a

ft

er a few seconds, 

water starts to 

fl

ow out of the outlet hose.  Switch on 

the RAD7 (have the printer switched on if you are 
using it. 

Th

e RAD7 will then print a header for the 

data printout, including a review of the setup, before 
it gives you a ‘Test’ prompt.) 

Provided that the RAD7 has been set up properly, see 
above, at the ‘

Test

’ prompt, push 

[ENTER]

 then the 

right arrow, to see ‘

Test Start

’ on the LCD, then 

push 

[ENTER]

 to start the test.

2.2  Speed of Response 

2.2.1  Measurement in Progress 

Th

e instrument is now measuring the radon in the 

water.  With high concentrations and short cycle 
times, and depending on the air and water 

fl

ow rates, 

it will take half an hour or more before there is much 
of a reading, and maybe 

fift

y minutes before you can 

rely on the count rate being close to the equilibrium 
value.  A

ft

er that you need to accumulate su

cient 

counts for the precision desired.  For example, 100 
counts would give a reading with a standard 
deviation of 10%.  At very low concentrations, it may 
take hours, and averaging over many cycles, to reach 
a su

ciently precise value. 

2.2.2  In

uencing Factors 

Th

ere are two processes requiring time.  One is for 

the air in the closed loop to approach equilibrium 
with the water and the other is for the RAD7 to 
respond to the changed radon concentration in the 
air loop.  

Th

fi

rst is primarily controlled by the water 

fl

ow rate and the second is determined by the half life 

of the 

fi

rst daughter of radon, namely 218-polonium.

2.2.3  Water 

ow rate 

At typical room temperature, the equilibrium 
coe

cient for radon in air and water is about 4:1.  

Th

at is the concentration of radon in air at 

equilibrium with water will be four times higher than 
the concentration in the water.  If the water were able 
to give up all its radon to the air it would take four 
times the air volume just to deliver the radon.  In 
practice the transfer is not complete so we may 
estimate that ten times the air volume is required. 

Considering the volume of the RAD7, the drying unit 
and the RAD AQUA, we can conservatively estimate 
the volume of the air loop to be of the order 4 litres.  

Th

erefore about 40 litres of water is needed to deliver 

the radon to the air loop before it can reach 
equilibrium.  A water 

fl

ow rate of V L/min will take at 

least 40/V minutes to deliver the radon. 

2.2.4  Air 

ow rate 

Th

ough important for thoron, see below, the air 

fl

ow 

rate is less critical to the radon response time.  For 
maximum speed of response, the air should keep 
circulating around the loop so that the air in the 
exchanger is continually being replenished with air 
from the measurement chamber of the RAD7.  

Th

us 

the shortfall from equilibrium and hence the 
e

ciency of transfer is maximized.  To achieve this 

the pump may be set to ON, see 1.4.1 and 1.5.2 
above. 

For a more relaxed operation, the pump may be set to 
AUTO, which will preserve the desiccant and 

Section 2 

 

Measurement Procedure

 10

Summary of Contents for RAD AQUA

Page 1: ...RAD AQUA Continuous Radon in Water Accessory for the RAD7 User Manual...

Page 2: ...rature Probe 8 1 3 Water Flow 8 1 3 1 Water Source 8 1 3 2 Water Level 8 1 4 Air Flow 8 1 4 1 Continuous Pumping 8 1 4 2 Pump on Auto 8 1 4 3 Mode set to Auto 8 1 5 Protocol 9 1 5 1 RAD7 Protocol 9 1...

Page 3: ...Exchanger to RAD7 method 2 14 4 4 Speed of Response 15 5 DRYSTIK 16 5 1 Passive DRYSTIK 16 5 2 Active DRYSTIK 16 5 3 E ect on Response Time 16 5 4 Custom designed Active DRYSTIK 16 6 BOATING 18 6 1 Re...

Page 4: ...ant Regeneration 20 7 5 RAD AQUA Troubleshooting 20 7 5 1 Rising Water Level 20 7 5 2 Spray Chamber Fills With Water 21 7 5 3 Air Path Integrity 21 7 5 4 Poor Spray Formation 21 7 5 5 Water Overflowin...

Page 5: ...temperature the coe cient is about 0 25 That means there is four times higher concentration of radon in the air than in the water so there is in e ect a gain of four times in the sensitivity of the s...

Page 6: ...plications where speed of response is a major goal WL1 is a compromise between the two 1 1 2 Temperature Probe The probe is inserted through the stem adapter A little petroleum jelly may help it to sl...

Page 7: ...nt left at this end to the air inlet filter with 1 8 ID tubing at the filter end which is then placed on the RAD7 INLET The Luer taper ensures an airtight connection Fig 2 RAD7 AQUA Standard Setup 1 2...

Page 8: ...ill be observed inside the body of the exchanger Water will accumulate inside the base and overflow out through the hose connector s If the flow rate is very high it may be necessary to utilize both h...

Page 9: ...ntrations to gain better statistics choose AUTO SETUP THORON Choose OFF SETUP PUMP Choose ON or AUTO depending on choice see above SETUP TONE Choose what you like SETUP FORMAT Choose what you like but...

Page 10: ...100 counts would give a reading with a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ciently precise value 2 2 2 In uencing Factor...

Page 11: ...QUA can be running with a low water flow rate and the RAD7 pump can be on AUTO also For fast response however it is essential to force the RAD7 to stay in SNIFF mode Setup Mode Sniff ENTER It will the...

Page 12: ...radon concentrations This water temperature data is obtained by exporting it from the temperature logger software see Section 3 1 5 below For specific instructions on downloading and graphing RAD7 dat...

Page 13: ...emperature data is stored in the temperature logger It is necessary to download the RAD7 data to a PC preferably using DURRIDGE s CAPTURE software and to download the temperature logger data using the...

Page 14: ...say 4 L min will mean a time delay for the thoron to reach the exchanger of no more than 2 5 seconds 4 3 2 Exchanger to RAD7 method 1 The RAD7 pump will typically generate an air flow rate of around 0...

Page 15: ...ponse of the RAD7 to thoron is limited not be the half life of the polonium daughter but by the time it takes to get the sample into the measurement chamber We have seen that with the highest sensitiv...

Page 16: ...7 will keep the air sample in the RAD7 below 7 RH It will also last for a very long time and add only a tiny volume to the air loop maintaining a fast response See Fig 3 5 3 E ect on Response Time Wit...

Page 17: ...Fig 3 RAD AQUA con guration with Active DRYSTIK and Water Switch Section 5 DRYSTIK 17...

Page 18: ...eded increases significantly The Rule 3700 delivers 6 L min through a W4 nozzle but it takes over 15A One 8 hour run would completely discharge a 120 AH battery 6 1 2 Minimizing T1 with reduced air vo...

Page 19: ...reducing the advantage of the three pumps to only 5 to 10 minutes But with the Rule 3700 the extra 20L to 40L of water would take only 3 or 4 minutes to deliver thus preserving almost all the 20 minu...

Page 20: ...ir sockets and push the lid onto its hinges The wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time The instrument should in any case be returned every year for...

Page 21: ...or leaks 7 5 3 Air Path Integrity When drawing a sample from a remote location air path integrity is essential to prevent dilution of the sample with ambient air Always make sure that there are no lo...

Page 22: ...ter inputs into river dominated areas important The Chao Phraya River Gulf of Thailand Limnology and Oceanography 51 2232 2247 Povinec P P P K Aggarwal A Aureli W C Burnett E A Kontar K M Kulkarni W S...

Page 23: ...2008 Determination of transport rates in the Yellow River Bohai Sea mixing zone via natural geochemical tracers Continental Shelf Research 28 19 2700 2707 167 Povinec P P H Bokuniewicz W C Burnett J C...

Page 24: ...67 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2018 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo are tra...

Reviews: