12
RMO200
Getting Started
Now there are two choices:
1.
Starting a new test with a different current value:
– press
the
STOP
button to change to the
Sing
menu,
–
set a new current value and/or another memory
position,
– confirm
your
entries by pressing the
ENTER
button,
–
start the test by pressing the
Ω
button.
2.
Starting a new test with the same current value:
– press
the
STOP
button to change to the
Sing
menu,
– press
the
ENTER
button to change to the
Test
menu,
– press
the
Ω
button to run a test.
The current value remains the same as in the
previous test, the memory position increases by one.
6.6 Viewing Results of Previous Tests
RMO200 stores up to 100 test results. They can be viewed using the
Memory
menu at memory positions 0 ... 99.
Selecting a memory position on the
Memory
menu displays:
• the current fed into the test object,
• the memory position of these results,
• the test object’s measured resistance,
• the date or the time,
of this particular test.
To view the creation date of the test rather than the time, position the cursor on the
latest number of the memory position (here
0
), and use the
RIGHT
button to toggle
between date and time.
6.7 Printing the Test Results
RMO200 stores up to 100 test results. They can be printed using the
menu at
memory positions 0 ... 99.
Connect the RS232 port of the RMO200 and the printer with a printer’s serial cable.
Go to
menu using
LEFT
and then
UP
button. Selecting first and last memory
position and press
ENTER
button. The results that are located in the range between
this memory locations, including the chosen memory locations, will be printed.
Figure 6-9:
The
Mem
menu showing the
previously saved test results
of memory position
30
Figure 6-10:
The
menu showing
first and last memory
position