5
RMO200 Functionality
4 Functionality
4.1 Single Test
The RMO200 generates a filtered (true) DC current and outputs it in an
automatically regulated current ramp. The current increases with a constant
slope from zero up to a set maximum value. When the current has reached its
maximum, RMO200 measures the resistance of the test object. After test the
output current ramps back down to zero with a constant slope. By sloping the
current up and down, magnetic transients are virtually eliminated.
4.2 Continuous Test
RMO200 can generate DC current continuously using the
Cont
menu. In this
menu the current can be chosen the same way like in the
Sing
menu, but the
duration of the test can be preset. The test is started pressing the
Ω
button.
During the test, a new result is shown on the display and stored into the PC
(RMOWin) approximately each second. Using RMOWin the result can be printed
like an Excel table which later can be shown as a diagram and used for a report.
Test is completed when the test time previously set in the menu has elapsed,
or when the
STOP
button is pressed.
By pushing
STOP
button test could be performed at any time. In case of test
interruption by pushing the
STOP
button, the last measured result will remain
saved in the internal memory.
Upon expiry of the selected test lasting time, the last result remains saved at
selected memory location.
Note 1:
Choose measuring current so that for the expected resistance you have
maximum value of it.
Note 2:
Using RMOWin enables
starting and controlling RMO200 from the PC.
The results could be stored to PC.
Using RMOWin after testings enables result transfer from the
internal RMO200 memory to
PC.
Note 3:
At the start of the test (Single and Continuous), a check of cables
connection is done. In case of disconnection, or erroneous polarity,
an alarm is activated, and the error message is shown on the display.
Voltage sense cables disconnection, occured during the test, will cause an
erroreous result. The display will not show this asan error message.
During tests the high continued DC current will heat up the test set. To
prevent RMO200 from overheating, duty cycles apply. The duty cycles
depend on the used test current, please refer to section "Duty Cycles".