Issue: 16.08.2010
Instruction Manual DVZ-SKA “BIOMASTER“-PLUS
Page 23 of 39
valve at the low point and drain water as required until the piping error can be
corrected.
3. If the above cheek out, detergent foam may be blocking the Treatment Tanks vent
(this is possible but unlikely, since the vent and traps are sized for this worst case).
4. If the back pressure persists, wash down the Treatment Tank vent from its upper
termination with a stream of high pressure water to clear any possible blockage.
Before doing this, set discharge Pump(s) to ON and watch Disinfecting tank sight
glass to prevent flooding.
5. If the Treatment Tanks vent is not the correct size, back pressure will almost certainly
occur every time a highsuds detergent is introduced into the system..
9.8 HIGH WATER LEVEL IN TREATMENT TANK
Obvious, check all valve positions, also in external piping to the overboard connection,
if this is found in order.
There can be several causes; The spillover hose (inside treatment tanks, to disinfectin
tank) are clogged, the vent on the disinfecting tank is blocked, or the Treatment Tanks are
overdue pumpout.
1. To clean the spillover, open inspection hatch on top of the Treatment Tanks and wash
out the vent with a stream of water. It is preferred to flush- and pump out the contact
tank after this action.
2. To clean the spillover inside the Treatment Tanks by flushing with water.
9.9 WATER SPILLING OUT OF THE CONTACT TANK VENT
This almost certainly is caused by failure of the Discharge pump, its controls, or by
discharge valves being closed.
1. Check valve positions, also in connecting pipe to overboard discharge.
2. Make sure the pump switch is set to Auto and that power is present.
3. Make sure the suction valve from the Disinfection Tank „V3“ is open
4. Check out the starter, motor, (especially thermal protection switch) and pump.
Controls should be set per "Set-Up for Operation" of this manual. Check function by
manual start of pump, if this does not work, check level switch for proper function, if
needed, clean out or replace.
Summary of Contents for Biomaster PLUS Series
Page 15: ......
Page 16: ......
Page 17: ......
Page 19: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 83: ...Issue 16 08 2010 Instruction Manual DVZ SKA BIOMASTER PLUS Page 33 of 39...
Page 89: ...Issue 16 08 2010 Instruction Manual DVZ SKA BIOMASTER PLUS Page 39 of 39...