Operation Mode
The system will always power on in default
Operation Mode
with all operational parameters set as
they were during the last use. True bottle speed will be displayed along with any alarm conditions.
Setting Bottle Speed
The user can change the bottle speed during operation by pressing the
Up
Button (top right) to
increase speed or the
Down
Button (bottom right) to decrease speed at any time. The
Up
button will
increment the bottle speed by 0.01 rpm with each press and the
Down
button will decrement the
bottle speed by 0.01 rpm each press. Holding either button down for about 5 seconds will allow for
faster display value changes when a larger numerical change is required.
The standard bottle speed range for this system is 0.23 rpm to 8.02 rpm (using 110 mm diameter
bottles).
Setup Mode
Pressing the
Select
button (top left) once enters the
Setup Mode
. The
Setup Mode
is used to set the
bottle diameter and direction of rotation. Once set, these parameters are usually not changed unless
different size bottles are used. Values set in the
Setup Mode
are not erased when power is removed
from the system.
To change a parameter value, press the
Select
button until the desired parameter is displayed. Press
the
Enter
button and the display will show the existing parameter value. Press or hold the
Up
button
to increase the parameter’s value. Press or hold the
Down
button to decrease the parameter’s value.
When satisfied with the value change, press the
Enter
button to save the new value to memory and
automatically return to
Operational Mode
.
To check a parameter value without changing it, press the
Select
button until the desired parameter
is displayed. Press the
Enter
button to display the existing value. Press the
Enter
button again to exit
the
Setup Mode
and return to normal
Operation Mode
.
NOTE:
If no buttons are pressed for 30 seconds, the system will automatically exit
Setup Mode
and
return to normal
Operation Mode
.
Bottle Diameter
The Standard Roller Apparatus system is designed to accurately display actual bottle rotation speed.
Once a bottle diameter has been programmed, the unit will automatically calculate and adjust itself
to roll the bottle correctly at its set speed. Different bottle manufacturers produce bottles of different
bottle diameters. It is not recommended that bottles of different manufacturers be mixed on the
same unit if an accurate bottle speed is to be obtained.
The Standard Roller Apparatus relies on the
maximum
bottle diameter for proper bottle speed
indications. Traction rings and any outside ridges on the bottle must be considered in calculating the
maximum bottle diameter. Carefully measure (in millimeters) the maximum diameter of a sample
bottle to be used. If traction rings or ridges are at both ends of the bottle, measure both ends and take
the average. Using the obtained measurement(s), enter the bottle diameter (in millimeters).