OPM BWT PERMAQ® Pico Rev. 2.0 EN
Technical data | 10
EN
53
10.5 Technical data for the BWT PERMAQ® Pico Duo 20/10–
90/80
20/10 30/20 50/40 60/50 70/60 80/70 90/80
Permeate:
Offline output, free run-off
1)
l/h
> 180
> 260
> 580
> 1150
> 1400
> 1700 > 2000
Offline output, free run-off
2)
l/h
> 200
> 300
> 650
> 1250
> 1500 > 1800 > 2200
Salt retention rate
%
99.0
99.5
Permeate yield (WCF)
%
80
80
Feed water
Pressure range
bar
2.5 – 6.0
Temperature range
°C
5 – 30
Hardness/with antiscalant dosing
°dH
°fH
0.0 / 25
0.0 / 45
Iron (Fe) and manganese (Mn)
mg/l
< 0.1
Salt content, total dissolved solids
(TDS)
mg/l
< 1,000
Silt density index (SDI)
%/min
< 3
Oxidants
mg/l
< 0.05
Silicate (SiO
2
)
mg/l
< 20
Hydraulic connections:
Feed water
Inch
3/4"
1"
Permeate
Inch
3/4"
1"
Concentrate
mm
13
mm / FT version 8
mm
13
mm / FT version
10
mm
General
Degree of protection
IP
63
Ambient temperature
°C
5 – 35
Electrical connection / fuse protec-
tion
V/Hz/
A
400/50/10
Electric power consumption
kW
2 x 0.85
1.6 +
2.3
2 x 2.3
2.3 +
3.1
2 x 3.1
2 x 3.1
Type of valve
VDC
24
Connector standard
CEE
16; 3PNE
1)
Feed water: TDS ≤ 1,000
mg/l; 15
°C; 2.5
bar
2)
Feed water: TDS ≤ 1,000
mg/l; 15
°C; 4.0
bar
Table
17:
Technical data for the BWT PERMAQ® Pico Duo 20/10–90/80
Pico Duo assembly connection set
Type 10–30
Type 40–60
Type 70/60
Part and identification no.
6-473224
6-473217
6-473218
Table
18:
Part and identification no. for BWT PERMAQ® Pico Duo connection set