background image

Printed in U.S.A. 3/19

FR# 444131-00 Rev. 2

©Copyright 2019 Dwyer Instruments, Inc.

DWYER INSTRUMENTS, INC.

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Phone: 219/879-8000
Fax: 219/872-9057

www.dwyer-inst.com
e-mail: info@dwyermail.com

CO2 Offset (OFC)

This value allows the CO

2

 Concentration to be adjusted by a fixed amount to match 

another calibrated measurement. The display shows the current CO

2

 concentration 

value plus any previous offset value.

Auto Serial Configuration (AUT)

This value enables or disables the automatic baud rate detection.  If the device fails 

to communicate on the MS/TP bus or the serial configuration is not 8 data bits, no 

parity and 1 stop bit, then this value should be set to “OFF”, and the serial configured 

manually.

Baud Rate (BAU)

This value provides the selection of the desired serial baud rate. This value is only 

visible when the value of AUT is “OFF”.

Parity Selection (PAR)

This value provides the selection of the desired serial parity. This value is only visible 

when the value of AUT is “OFF”. 

Stop Bits Selection (STP)

This value provides the selection of the desired serial stop bits. This value is only 

visible when he value of AUT is “OFF”

Reset To Factory Defaults (RST)

This value, when set to “YES”, will reset all user settings to their default values and 

reset the device. This applies to all settings including BACnet writable settings.

Calibrating Sensor

Step 1:  Remove the cover as shown in Figure 1. 

Step 2:  Remove one of the gas nipple covers on the CO

2

 sensor and attach tubing 

          from the gas pressure regulator to the nipple (See Figure 5).

Step 3:  Attach the terminal block accessory to the circuit board so that the power wires  

          line up with terminals 1 and 2. Plug in the power supply to power up the 

     transmitter.

Step 4:  Hold housing so that the sensor is in the vertical plane as shown in Figure 5.

Step 5:  Follow the steps in the accessing parameter section to access the calibration 

          parameter (CAL).

Step 6:  Press and hold the Up and Down arrows for 3 seconds.

Step 7:  Flow zero reference gas at 0.3 SLPM for 5 minutes.

Step 8:  Press and hold the Down button for 0.5 seconds.

Step 9:  Flow the full scale reference gas at 0.3 SLPM for 5 minutes.

Step 10: Press and hold the Up button for 0.5 seconds.

Step 11: Exit the parameter menu.

Step 12: Disconnect the power supply from the power source and remove the terminal 

     block from the circuit board.

Step 13: Remove tubing from sensor and re-attach the gas nipple cover to the sensor.

Step 14: Re-attach the cover to the back plate.

MAINTENANCE/REPAIR

Upon final installation of the Series CDTA, no routine maintenance is required. The 

Series CDTA is not field serviceable and should be returned if repair is needed. Field 

repair should not be attempted and may void warranty.

WARRANTY/RETURN

Refer to “Terms and Conditions of Sale” in our catalog and on our website. Contact 

customer service to receive a Return Goods Authorization number before shipping the 

product back for repair. Be sure to include a brief description of the problem plus any 

additional application notes.

Default Value Minimum Value

Maximum Value Increment

0 PPM

-500 PPM

500 PPM

1 PPM

Setting Value Description

“ON”

“OFF

Auto baud enabled, assumes 8 data bits, no parity and 1 stop bit

Auto baud disabled, serial baud rate, parity, and stop bits must be 

set manually

Setting Value Baud Rate (BPS)

9.6 K

19.2 K

38.4 K

57.6 K

76.8 K

115.2 K

9600

19,200

38,400

57,600

76,800

115,200

Setting Value Description

“NON”

“EVE”

“ODD”

No Parity

Even Parity

Odd Parity

Setting Value Description

1

2

One Stop Bit

Two Stop Bits

Figure 9: Calibration

European

North American

This symbol indicates waste electrical products should not be disposed 

of with household waste. Please recycle where facilities exist. Check with 

your Local Authority or retailer for recycling advice.

Summary of Contents for CDTA-2E000

Page 1: ...ength NDIR Humidity Capacitive polymer Temperature Solid state band gap Range CO2 0 to 2000 or 5000 PPM CO2 depending on model Humidity 0 to 100 RH Temperature 32 to 122 F 0 to 50 C Accuracy CO2 40 ppm 3 of reading RH 2 10 to 90 RH Temperature 1 C 25 C Temperature Dependence CO2 8 ppm C at 1100 ppm Non Linearity CO2 16 ppm Pressure Dependence CO2 0 13 of reading per mm of Hg Response Time CO2 2 mi...

Page 2: ... two 6 M2C type screws provided 4 Pull wires through sub base hole and make necessary connections 5 Reattach cover to base plate WIRING Figure 2 shows how to connect the CDTA in a network containing a common power supply Use a cable containing two twisted pairs One pair is to be used for B and A The other pair is to be used for power and common This configuration is not suitable for AC supplies Us...

Page 3: ...oint temperature will be set to the new set point low limit temperature Set Point High Limit Temperature SOH This value sets an upper limit on the current set point value The set point high limit value is numerical setting that supports continuous change increment decrement by press and holding of either UP or DOWN button If the new set point high limit temperature is lower than the current set po...

Page 4: ...e vertical plane as shown in Figure 5 Step 5 Follow the steps in the accessing parameter section to access the calibration parameter CAL Step 6 Press and hold the Up and Down arrows for 3 seconds Step 7 Flow zero reference gas at 0 3 SLPM for 5 minutes Step 8 Press and hold the Down button for 0 5 seconds Step 9 Flow the full scale reference gas at 0 3 SLPM for 5 minutes Step 10 Press and hold the...

Reviews: