M24

 OWNERS MANUAL 

 

IMPORTANT WARNING 

BEFORE USING PROFESSIONAL MODEL M24, 
READ ALL INSTRUCTIONS AND  SAFETY DATA 
AT THE END OF THE DOCUMENT. 

 

This manual covers the operation of the M24 Electronic Gold Tester.

     

 

RS MIZAR, INC. 

has been issued two Patents for its development on 

the M24 full range Gold Tester.  This new process gives accurate and non-
destructive test results all the way to 24K gold.

  

It can help determine Gold 

Karat value in as little as 2 seconds. 
 
This process can distinguish non-gold and gold plate from true “Karat Gold” 
without the destructive Fire Assay testing, which would change the value of 
the jewelry. 
 
RS Mizar now brings you the fastest and easiest Gold Analyzer for all the 
trades where jewelry and “Karat Gold” have to be quickly and accurately 
evaluated.  
 
The M24 Gold Tester: U.S. PATENTS 5080766, 5128016 
 

OTHER U.S. AND INTERNATIONAL PATENT RIGHTS PENDING 

 

 

 
 
 

 1 

 

 

 

CONTENTS OF PACKAGE 

 
 
 

 

1. M24 ELECTRONIC GOLD TESTER                                  
2. ERASER 
3. TESTING SOLUTION 

[No. 1995]

                  

4. OWNERS’ MANUAL 
5. ERASER PENCIL 

 

                

 
 
 

 

OPERATING INSTRUCTIONS

   

 

OPERATING INSTRUCTIONS FOR TESTING GOLD FROM 20K TO 24K 

When testing Gold from 9K to18K, go directly to instructions below for 9 to 
18 Karat testing.  When testing Gold from 20K to 24K, a buildup of test 
solution on the Reference Gold will have to be removed to obtain the proper 
results. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 3 

 

 

 
 

ITEM 5    When “TESTING” Light goes out, remove Test Gold.  Read LED 
Display of Karat Value.  Clean Test Gold thoroughly with soft paper towel. 
Thoroughly wipe the Test Solution from the Test Well using a paper towel 
followed by a Q-Tip.  If re-testing is desired, see “DO’S & DON’TS”, items 
“8” and “11”. 
 
ITEM 6    Check to be sure that there is not excessive air in the bottle and that 
the cap is tightened firmly.   
 

PRECAUTIONS 

 

After testing Gold item, it should be cleaned with a buffing wheel.  
However, it must be cleaned with the gray eraser before any new test 
because of the residue left on the Gold from the previous test. 

 

Improper of insufficient cleaning of the test Gold before testing can 
increase the karat value readings. 

 

Your unit is operating correctly if occasionally you do get a reading in the 
next range.  This can happen when the karat value of the Gold you are 
testing is very close to the odd karat value, i.e. 17K, 19K, etc.  Your unit is 
operating correctly when this occurs and indicates the Gold you are testing 
is very near the in-between value. 

 

The excess air must be squeezed out of the test solution bottle between 
tests or when stored.  Squeezing out the excess air keeps the solution 
fresh.  When all of the air cannot be squeezed out it needs to be replaced.  
In any case, the bottle must be replaced 6 months after first opening!  “Un-
fresh” test solution can cause the karat readings to go lower by one range.  
You should always have a fresh bottle of test solution available.  Record 
here the date the bottle was opened:______________________

  

                                                                              
 
 
 
 
 
 
 
 

       

 5 

 

 

 
 

SPECIFICATIONS 

 

M24 GOLD TESTER 
M24 Range: 9, 10, 12, 14, 16, 18, 20, 22, and 24 Karat 
 
Automatic Shutoff after each test: Approximately 14 seconds 
 
Test times:  2 seconds from 9-14 Karats 
 

            2-15 seconds from 16-24 Karats 

 
Delay before beginning test: 12 seconds 
 
Display time: 12 seconds 
 
Gold colors: Yellow, White, Red, Green (see “Gold Composition”) 
 
Non-destructive testing 
 
Distinguishes Gold Wash and Flash from Solid Gold 

  Can be used to 

distinguish Gold Plate from Solid Gold (see “Testing for Gold Plate”) 
 
Dimensions:   
Length ....................................................................................... 6.3” 
Width..........................................................................................4.0”      
Depth..........................................................................................1.5” 
Weight: With Battery..................................................................8.0 oz. 
Operating Temperature:  65

°

 F to 90

°

 F (19

°

 C to 32

°

 C) 

Battery:  One (1) standard 9 Volt 

 
 
 
 
 

 2 

 

 

 

STEP A - Sharpen the Pencil Eraser and then cut off the end with a sharp knife or razor 
so that it is flat and round (about .160” or 5/32”). Then place the eraser into the Test Well 
and firmly rotate back and forth for about four seconds.  Then brush debris from the Test 
Well.  This preparation will eliminate the excess buildup on the Reference Gold.  For all 
tests of 20K to 24K Gold do this step of removing excess buildup on the Reference Gold 
with the sharpened and blade-cut Pencil Eraser.  A dirty Eraser tip will not sufficiently 
clean Test Well. 
STEP B - Go to “item 1” below. 
 

OPERATING INSTRUCTIONS FOR TESTING GOLD FROM 9K TO 18K. 
 

ITEM 1    Thoroughly clean immersible area of Test Gold with the Gray Eraser (see 
“DO’S & DON’TS”, item 5.) Attach Alligator Clip to Test Gold away from immersible 
area. 
 
ITEM 2    Loosen the cap of the Test Solution and have a soft paper towel and some Q-
TIPS handy. 
 
ITEM 3    Press “TEST YELLOW GOLD” or “TEST WHITE GOLD” (to match the type 
of Gold that you are testing) and within 

two seconds, 

put three or more drops of Test 

Solution into the Test Well.  Quickly squeeze all of the air out of the Test Solution Bottle 
and tighten cap firmly.  (See “item 6”) 
 
ITEM 4    Observe that one of the Karat Value Lights is on.  Any Karat Light may be lit; 
this is determined by a previous test or length of time that the unit has been off.  After the 
Light goes out, (within one second), carefully immerse enough of the Test Gold into the 
liquid to start the “Testing” Light being careful 

not

  to touch the reference Gold at the 

bottom of the Test Well and being careful 

not

 to touch the Alligator Clip to the liquid.  

Hold the Test Gold without moving it until “TESTING” Light goes off.  This will take 2 
to 15 seconds, depending on the Karat value of the Test Gold.  User should brace the 
“heel” of his hand on the unit or on the same surface as the unit rests on.  When possible, 
brace the Test Gold on the inside of the Test Well just under the surface of the liquid and 
be very careful not to touch the Reference Gold at the bottom of the Test Well. 
It is important that the Test Gold not be moved about in the Test Solution until a reading 
is obtained.  

 4 

 

DO’S & DON’TS 

1)

 Don’t use Testing Solution for more than one test; errors will result. 

2) 

Don’t touch Test Gold or Alligator Clip lead to Reference Gold at bottom of Test 

Well.  If this happens remove liquid and re-do test procedure. 

3) 

Don’t let Alligator Clip touch liquid. 

4) 

Don’t leave liquid in test well.  This will void your warranty. 

5) 

Do clean Test Gold before testing if it isn’t bright and shiny. Unclean Gold may 

result in inaccurate readings. 

6) 

Do keep Gold chains very tight (link to link) during testing. Use the fewest number 

of links (one if possible) between the Alligator Clip lead and the Gold link to be 
immersed in to the liquid. Three problems can exist with testing chains: 
a. There may be dirt or contamination between links that would require intensive 
cleaning (i.e. ultrasonic) before testing. 
b.   Some chains, especially certain European manufactured chains have a thin 
non-metallic coating that must be removed from an area before testing.  The 
eraser pencil may be used to remove this coating. 
c.   When testing chains, remember that the solder often used at the link(s) is of a 
lower Karat than the rest of the chain.  Therefore, avoid testing in that area. 

7) 

Do make sure good contact is achieved between the Test Gold and the Alligator 

Clip lead.

  

If the object to be tested is too large to fit in the clip lead, simply hold 

the clip lead lightly against the Test Gold during test. 

8)

 Do always clean the Test Gold with the Gray Eraser before retesting in the same 

location.  Pink eraser can be used but it requires considerably more pressure and 
effort for accurate readings. 

9) 

 Do replace the Testing Solution every 6 months after opening. 

10)

Do use only within Operating Temperature 65

°

F-90

°

F. 

 

NOTES 

1) In the event the Alligator Clip contacts the testing solution, the Alligator 

Clip must be rinsed thoroughly with water and dried before using again. 

2) WARNING: Eye Hazard -- Read warning label on bottle of Test Solution.

 

3) 

IMPORTANT:

 The Testing solution is sensitive to exposure to air. Minimize the 

time cap is loose on bottle. 

 6 

 

 

 