background image

 

 
 
PREPARATION PRIOR TO TESTING A STONE 
 

 

The presence of dust, dirt, film and moisture on a stone may produce an inaccurate indication. Therefore, it is 
prudent to routinely clean the surface of each stone to be tested by using a damp soft clean fabric followed by 
drying using a dry soft clean fabric.   

 

 

Remove the protective cover from the probe. 

 

The  tip  of  the  probe  should  be  cleaned  frequently  by  wiping  the  tip  across  the  surface  of  a  piece  of  bonded 
paper or a business card.  Copier paper and fax paper are to be avoided. 

 

The  tip  of  the  probe  may  be  damaged  by  accidental  drop.    Always  place  the  protective  cover  over  the  probe 
immediately after use.  Probe tip damage is exempted from the warranty. 

 

During testing the users hands should be clean and dry.  Do not wear gloves. 

 

The stone to be tested should be allowed to reach room temperature. 

 

For maximum accuracy, minimize handling of the stone prior to testing.  If the item to be tested consists of a 
setting  containing  many  stones  requiring  more  than  ten  seconds  to  test  user  must  avoid  heating  the  setting 
with his hand; allow the item to be untouched for about fifteen seconds. 

 

Loose stones may be placed on the testing platform (9) at any time and be tested immediately. 

 

Best results are obtained by performing tests in still air.  Avoid fans, drafts or air movement from  air      
conditioning or heating sources.  Avoid breathing directly on the probe tip while testing. 
 

“AUTO” TESTING 
 
“Auto”-Testing can be performed when the temperature is within 70-88 F and the stone size is between .08 and .75 
Karats.  For all other conditions use the manual mode and adjust to the calibration chart (p.3).  The automatic testing 
mode (“Auto”-Testing) is achieved by rotating the thumbwheel counter-clockwise until a “click” is felt. 
 

1. 

Perform the operational checkout as described above. 

2. 

The twelve lamp Auto/Manual Range Indicator  is segmented as follows:  
 

   

 

 

 

 

 

3. 

The ambient temperature should be between 70

F and 88

F for “Auto” Testing.  Normally Lamp #3 will be 

lit. 

4. 

Hold the diamond tester in your hand so that your fingers are in contact with the conductive panel (6).    

5. 

With the other hand pick up the stone to be tested by its setting or place a loose stone in the appropriate 
well in the testing platform held with that hand.  

6. 

Guide the probe tip to the center of a facet on the stone. 

7. 

Touch  the  probe  tip  to  the  stone  so  that  the  probe  tip  is  firmly  and  squarely  against  the  facet  of  the 
stone.  Excessive force which might bend or break the tip must be avoided.   

8. 

Make sure the probe is perpendicular to facet of stone while testing. 

9. 

After each test the display should return to the original lamp. 

10. 

As  always  some  stones  require  further  testing  to  determine  what  they  are.  In  the  case  of  some 
Moissanite Stones especially the 

“Forever One” Moissanite stone try using UV lignt on the stone while 

testing will help like the ones used for counterfeit money e.t.c. 

 

                                                                  

 

 
 

 

 

Congratulations on your purchase of the Prestige Series II, 
the world’s finest diamond tester.  With this unique 
Instrument you will obtain the most information about the 
stones you are testing.  You also will have the versatility 
of  selecting  between 

the  “auto”  mode  with  valuable 

sensitivity  or  the  “manual”  mode  which  allows  for 
adjustments for extremes in temperature or stone size. 
 
IDENTIFICATION OF PARTS 
 1.  Probe 
 2.  Range Indicator 
 3.  Auto/Manual Range Adj. 
 4.  Wait/Low Battery Lamp 
 5.  Battery Compartment 
 6.  Conductive Panel 
 7.  Testing Platform for Loose Stones 
 

8.  “ON” Button 

 

9.  “OFF” Button 

 

*PRIOR  TO  TESTING,  IT  IS  RECOMMENDED  THAT  THE 
TIP  OF  THE  PROBE  BE  CLEANED  BY  GENTLY  WIPING 
ACROSS A CLEAN SHEET OF PAPER * 

 
OPERATIONAL CHECKOUT 
1.  Rotate  the  Thumbwheel  (3)  in  a  counter-clockwise 

direction beyond the “click”. 

2.   Press the “ON” button (8). 
3.   The Wait Lamp (4) will be lit. 
4.    All  other  display  lamps  will  remain  extinguished  for 

the  wait  period,  which  is  typically  less  than  nine 
seconds at  room  temperature.   The  time can  increase 
with  low  temperatures  or  a  weak  battery.  After  the       
wait period there is a single beep. 

5.  If the waiting time exceeds twenty seconds or the Wait 

Lamp fails to extinguish replace the battery. 

6. The Diamond Tester is ready to use after the Wait Lamp 

extinguishes and a display lamp (#1 to #8) is lit. In the 
“Auto”  mode  display  lamp  3  is  normally  lit  after  the 
unit has warmed up, typically ten to twenty seconds. 

7.    If  the  ambient  temperature  is  not  within  the  range  of 

the automatic circuitry (70-88 F, 21-31 C) the diamond 
tester  is  still  able  to  test  a  stone  by  testing  with  the 
Manual Range Adjust (3). 

8.    When  testing  is  completed  press  the  “OFF”  button.  

Always  turn  off  the  unit  after  testing  to  increase 
battery life. 

  
 

“AUTO-SET” 

This  adjustment  affects  the  sensitivity  of  the  “automatic 
range” only.  It is pre-set at the factory to lamp #3.  If you 
prefer  a  higher  sensitivity  for  diamonds  you  may  adjust 
the  “auto-set”  (through  the  bottom  of  the  front  label)  to 
lamp  #4.   To  make  this  adjustment,  turn  the  thumbwheel 
(3) fully counter-clockwise until clicked.  Turn the unit on 
and  wait  twenty  seconds  to  allow  the  probe  to  reach 
maximum temperature.  Then with a jeweler’s screwdriver 
adjust  the  “Auto-Set”  to  “the  center  of”  lamp  #4.    If  you 
prefer a lower sensitivity follow the above procedure and 
adjust  to  “the  center  of”  lamp  #2.      This  adjustment  will 
not affect the Manual mode. 

 

AUTOMATIC SHUT-OFF 
The Diamond Tester will automatically shut off after 90 
seconds of inactivity.  Whenever the “On” button is 
pressed the timer resets to 90 seconds.  If the unit is shut 
off before you have completed your test simply press the 
on button and continue your testing as soon as the 
display comes on.  Always 

press the “OFF” button after 

testing to prolong battery life.

 

Reviews: