background image

  

 

 

 

  

 

All rights reserved. Reproduction or issue to third parties in any form whatsoever is not permitted without written authority from the proprietors.

 

 

 

Table of Contents 

 

1) Introduction

 .......................................................................................................................................... 4

 

2) General Information

 ............................................................................................................................. 4

 

2.1) Unpacking and Inspection

 ............................................................................................................. 4

 

2.2) System Components

 ...................................................................................................................... 4

 

2.3) Specifications

 ................................................................................................................................ 5

 

2.4) Outline Drawing

 ............................................................................................................................ 6

 

3) Operation and Installation

 .................................................................................................................... 7

 

3.1) General

 .......................................................................................................................................... 7

 

4) Sensor Static Calibration Verification

 ................................................................................................. 7

 

5) Declaration of Conformity

 ................................................................................................................... 8

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

 

Summary of Contents for DYN-C-1000-LN

Page 1: ...Model DYN C 1000 LN 2g 5g 10g 30g 50g 100g 200g Product Manual ...

Page 2: ...y Copyright All copyrights of this manual belonging to Dynalabs products are reserved It cannot be reproduced without written consent Disclaimer Dynalabs Ltd provides this publication as is without warranty of any kind express or implied including but not limited to the implied warranties of merchantability or fitness for a particular purpose This document is subject to change without notice and s...

Page 3: ...tten authority from the proprietors 3 Table of Contents 1 Introduction 4 2 General Information 4 2 1 Unpacking and Inspection 4 2 2 System Components 4 2 3 Specifications 5 2 4 Outline Drawing 6 3 Operation and Installation 7 3 1 General 7 4 Sensor Static Calibration Verification 7 5 Declaration of Conformity 8 ...

Page 4: ...minum housing with protection class IP68 Steel housing is also possible Dynalabs LN accelerometers provide Low noise high resolution with an outstanding noise performance from 9 to 680 μg Hz These accelerometers provide a wide frequency range 5 from 250 Hz to 1 500 Hz DYN C 1000 LN sensors offer the following options Custom Cable Length Custom Housing Material Custom Connector Base plate 2 General...

Page 5: ...mV g 1 350 540 270 90 54 27 13 5 Frequency range 5 Hz 700 1 150 2 000 2 300 2 700 2 900 2 500 Non linearity full scale 0 15 0 15 0 15 0 15 0 15 0 15 0 15 Noise in band μg Hz 9 21 40 100 180 340 680 Bias temperature mg C 0 2 0 5 1 3 5 10 20 Shock survivability g 2 500 2 500 2 500 3 000 3 000 3 000 3 000 Environmental Table 2 Environmental Specifications datasheet Protection Level IP 68 Operating Vo...

Page 6: ...ble 3 Physical Specifications datasheet Sensing Element MEMS Capacitive Housing Material Aluminum or Steel Connector Optional D Sub 9 or 15 pin Lemo Binder Mounting Adhesive or screw mount Base plate Optional Aluminum or Steel Weight without cable 10 g aluminum 20 g steel 2 4 Outline Drawing The dimensional properties of DYN C 1000 LN sensors are given below ...

Page 7: ...ower supply to the power ground Always use a clean power source and check the voltage range 4 Sensor Static Calibration Verification Using gravity voltage values are measured in the and gravity directions providing a value of 1 g The measurement should be made as follows When the sensitivity value of 1000LN series sensors is used with the data acquisition system the sensor shows 1 g with the effec...

Page 8: ...ted according to following EC directives 2014 35 EU Low Voltage Directive LVD 2006 42 EU Machinery Safety Directive 2015 863 EU RoHS Directive Applied standards EN 61010 1 2010 EN ISO 12100 2010 MIL STD 810 H 2019 Test Methods 501 7 High Temperature 502 7 Low Temperature 514 8 Vibration 516 8 Shock DYNALABS MÜHENDİSLİK SANAYİ TİCARET LİMİTED ŞİRKETİ declares that above mentioned products meet all ...

Reviews: