SYSTEM MANUAL Version 6.03i
85
18. Maintenance
General Care
Whilst the C-SCAN system has been designed for daily use it must be remembered that it
is a precision instrument and should be treated with a degree of care. When not in use the
Detector and Signal Generator should be stored in their soft cases and particular care
should be taken in the field with regard to the damage that can be caused by vehicles
when equipment is placed on the ground.
Battery Care
Both the Detector and Signal Generator are fitted with rechargeable batteries. These
should not be allowed to run completely flat as this can shorten the life of the battery.
When storing either the Detector or Signal Generator for long periods the batteries should
be isolated using the physical isolator switch on both systems. For prolonged periods of
storage the batteries should given a charge every 6 months.
When replacing either the rechargeable or alkaline batteries, only good quality batteries
should be used. Cheap batteries have inferior performance and are prone to leakage
which could damage the C-SCAN system.
Routine Checks
The C-SCAN system requires little-to-no routine maintenance. A general visual inspection
of the system should be carried out routinely to inspect for any physical damage, which
could indicate damage within the system that cannot be seen. Periodically check the
fittings and bolts to ensure they have not worked loose, which could lead to water ingress.
Signal Generator Battery Replacement
As every battery has a life expectancy it is likely that at some point the Signal Generator
battery will require replacement. Although Dynalog would recommend returning the Signal
Generator for battery replacement, it has been designed so that users with the
competency to carry out this task can do so with ease.
NOTE
: User replacement of battery while system is under warranty will invalidate any
warranty.
To gain access to the internal battery, ensure the
isolator switch is in the OFF position then remove the
8 bolts on the sides of the Signal Generator. Do not
remove any of the bolts on the top of the Signal
Generator. Once the bolts have been removed,
carefully lift the entire top plate (heatsink) away from
the casing, in an upward-left motion; be careful not to
damage any cables during this process. Once the
top plate is away from the casing, the cable
assembly can be removed through the access panel
on the underside and the top plate can be removed
completely and stored safely.
Using the strap fitted, lift the battery to an upright position so that it is possible to access
the terminals. Remove the terminal bolts and extract the battery completely. Only replace
the battery with a genuine Yuasa battery of the same type and specification. Any other
battery fitted will invalidate any warranty that may be in place.
Re-fitting is a reversal of removal.
8 x side bolts