medICal deVICe rePorTIng requIremenTs
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL
101
Reporting any Incident of Patient Discomfort
Dynatronics recommends that if discomfort of any level is reported by the patient, the treatment be stopped immediately.
The device and all accessories in use during that treatment should be isolated and held for inspection. Make a note of
treatment parameters that were in use during the treatment including intensity settings. Also note environmental factors
that were observed during the treatment (office lights flickering, static electricity discharge, other devices in use on the same
power source or in the same room, etc.)
The incident should be reported immediately to Dynatronics Customer Service at 1-800-874-6251. The customer service
representative will inform you if it is necessary to send the device and/or accessories to Dynatronics for inspection.