hIgh VolT InsTruCTIons
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL
43
CUSTOM (Duty) CYCLE TIME SELECTIONS
• Press the FUNCTION key located on right-side of the device faceplate.
• Using the CONT/REST toggle key, select CUSTOM in the CONT/REST window.
• Using the MODE toggle key, select CUST CR in the MODE window. Selections will be illuminated GREEN.
• Custom Contraction and Custom Rest cycle times may now be set by using the Up and Down arrow keys
located next to Custom Contraction and Custom Rest windows. Available ranges for Contraction (ON) times
are 1-120 seconds, for the REST (OFF) times 1 to 300 seconds. Remember, the REST time cannot be less than the
Contraction time. Pressing and holding the START key until the beep is heard will save the current Contraction/
Rest settings as the default.
6. Choose the RAMP setting.
FUNCTION KEY: To access the RAMP settings, press the FUNCTION key.
Press the FUNCTION key on the right side of the face plate. Press the RAMP toggle key to select RAMP time.
Selections include: 0.5, 1.0, 1.5, and 2.0. Ramp time is applied to both the start and end of the contraction. Ramp
time is in addition to the contraction time itself. A ramp setting is not applied to the Continuous duty cycle except
when using a probe. With a probe treatment set to a Continuous duty cycle, a ramp time of 3 seconds is automatically
applied.
7. INTENSITY.
Using the Up and Down arrow keys to the right-side of the INTENSITY display, set the Intensity. Remember, when
you increase intensity, current to the patient begins. Therefore, START should be pressed immediately after setting
the intensity to begin the treatment timer.
8. Press START.
Press START. When you press start, the treatment timer begins counting down.
NOTE: Never use High Volt to treat any conditions which contribute to loss of sensation, or an area where the
patient cannot feel the electrical stimulation.
NOTE: SAVING DEFAULTS. If the treatment you have just set up is the most common High Volt setup you use,
the treatment parameters can be saved as the defaults for your own device. After setting up the treatment, press
and hold the START key for two seconds. At the end of two seconds, a beep will sound indicating the treatment
parameters have been saved. The next time this modality is chosen, the parameters will be selected automatically.