CombInaTIon TheraPy InsTruCTIons
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL
65
Combination Therapy Setup
1. Press the COMBO key on the 825 and 925. SOUND will be illuminated GREEN in the CHANNELS window and
an automatically selected STIM channel will be illuminated in YELLOW. Other available channels not in use will be
illuminated in WHITE.
2. Plug the combo lead wire into the active STIM channel jack of the 825 or 925. The banana connector end plugs
into the COMBO input jack on the right-side of the unit. The pin end attaches to the dispersive electrode.
3. Apply the dispersive electrode to the patient
4. With SOUND illuminated GREEN, select parameters for the Ultrasound portion of the COMBO treatment
following the instructions found in the manual.
5. Press START.
6. When using the 825 or 925 press the single-channel STIM option of your choice. The selected STIM channel’s
illumination will change from YELLOW to GREEN and become the focus treatment. NOTE: If selecting a Stim channel,
other than the default channel, the illuminated channel will change from WHITE to GREEN. The SOUND channel will
now be illuminated YELLOW. Proceed to set up the STIM treatment following the modality instructions found in the
manual.
Remember, High Volt must use the designated HI-VOLT Channel. The system defaults to a PREMOD treatment
selection unless another default therapy has been selected previously.
7. Apply the conductive gel to the Ultrasound treatment site now. Place the soundhead at the treatment site making
good contact with the skin.
8. Raise the Stim intensity. KEEP THE STIM INTENSITY LOW! If the soundhead is in proper position and coupling
is good, the patient will feel the current. If the patient does not feel the current, check to be sure coupling is good and
make sure you have used ample conductive gel.
9. Press START. Both STIM and ULTRASOUND will be activated.
10. STOP.
When the treatment time has elapsed, the therapy to the patient stops and a tone sounds notifying you of the
treatment end. Treatments in progress may be stopped at any time using one of the following methods.
NOTE: Although parameters and settings are selected for ultrasound therapy, no ultrasound power will be delivered
until you press START after both the ULTRASOUND and STIM modalities have been set up.