sImulTaneous TreaTmenTs
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL
67
Simultaneous
Treatments
The Dynatron 25 Series allows many combinations of simultaneous treatments to be delivered at once using available
channels. Simultaneous treatments are not the same as COMBO treatments. A COMBO treatment combines Ultrasound
with a Stim therapy into a single treatment. A COMBO treatment is always delivered to one patient. Simultaneous treatments
are independent treatments that are set up separately, that have separate treatment timers, and which may be delivered to one
or more patients at the same time.
There are very few limitations to the simultaneous treatments that may be set up with the Dynatron 25 Series. You can set up
any number of separate treatments as described below with the exceptions noted:
• Channels 1 through 4 may be used for any number of Interferential, Premodulated, Russian, or Biphasic treatments.
For treatments using one channel, the device will select the next available channel. For treatments using two channels,
the device will select a channel pair (1-2 or 3-4). Note: Channels 3-4 are available on 25 Series 625 and 925 only.
Set Up A Second Treatment
To set up the second (or third) treatment, after you have set up and started the first modality, press the modality key for the
second treatment to be set up. The device automatically selects the treatment channel(s) to be used. The GREEN channel
light shows you the channel(s) selected for this treatment. Plug the lead or cable into the corresponding output jack(s) before
you proceed with setting up the treatment. Select the treatment parameters for the second treatment following the setup
instructions for that modality provided in this manual. When parameters have been entered, press START.
Modify Simultaneous Treatments
You may VIEW and MODIFY parameters for the treatment channel(s) that is illuminated GREEN. While two or more
treatments may be in progress at once, the TIME / INTENSITY displays can show only the settings for the focus treatment
whose operational channels are illuminated GREEN. Other channels in use at that time will be illuminated YELLOW
indicating that the channels are active (delivering current) but their parameters are not currently displayed or in focus.