baTTery oPeraTIon
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL
78
Battery Operation
All 25 Series devices are manufactured with battery capabilities allowing you to deliver treatments wherever power may be
unavailable or unreliable. To use the optional battery, do the following:
1. It is recommended that a battery be charged for 24 hours prior to operating the 25 Series device. DISCONNECT the
battery charging cable from the battery while it is in use for treatment.
2. Plug the battery adapter into the jack labeled “BAT-INPUT 12V-DC” on the back of the Dynatron 25 Series console.
3. Set up and deliver treatments.
4. When battery power is reduced to approximately 11 volts, a low battery warning will be displayed in the lower
right-hand corner of the Treatment Display Screen “CAUTION: BATTERY LEVEL GETTING LOW!” The treatment
can continue however there will not be enough power to set up and deliver another treatment when the current
treatment has ended.
5. When the available battery power becomes too low to continue operating the device, the following message will
appear: “ERROR: BATTERY LEVEL TOO LOW FOR TREATMENT OPERATION. CHARGE IMMEDIATELY!”
The treatment intensity will ramp down, any treatments that were running at the time will stop, and the device will
shut down. Before battery operation can continue, the battery must be recharged.
Use ONLY a Dynatronics’ Approved Battery
Before purchasing or using an existing battery with a 25 Series device, contact Dynatronics or your Dynatronics
Representative to obtain specifications for a battery that may be safely used with a 25 Series device.
Only use a battery that CANNOT be recharged while it is in use. Disconnect the battery charger from the AC
power source before using the battery to supply power to this device.