dynaTron® TherMosTIM™ Probe
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
100
Using Stim With The ThermoStim Probe
Premod Treatments Setup
A Premodulated treatment uses one channel and two electrodes.
1. Select Channel
Using the Channels toggle key located beneath the Channels Window, select the channel into which the stereo plug
(F) has been inserted. Channels 1,2 ,3, or 4 will become illuminated when selected.
2. Time
Set the treatment TIME. Default Treatment Time is 10 minutes.
3. Select RANGE
HIGH, LOW, ALT (high/low alternating), CONSEC ( ½ high, ½ low), or FIXED (select 4,000 to 10,000 Hz).
4. Intensity
With the ThermoStim Probe in contact with the patient’s skin, increase INTENSITY (patient will feel the current)
using the arrow keys located next to the INTENSITY display. Do not increase INTENSITY beyond the patient’s
comfort level.
5. Press START
The selected electrotherapy modality parameters will be displayed and Time will begin counting down.
TREATMENT NOTES
• Apply sufficient pressure so that at least 50% of the probe face will remain in contact with the patient’s skin at all
times during treatment.
• All Contraindications, Warnings, and Precautions for Stim modalities apply when using the ThermoStim probe
• The ThermoStim Probe cannot be submerged in water
6. Press STOP
To stop a ThermoStim treatment prior to its completion, press and hold the FUNCTION KEY and press the STOP
key. Pressing STOP alone stops all treatments operating on the device. Using the TIME arrow keys to bring a
treatment time to zero will also stop a treatment.
7. Moving Between Stim and Thermal Displays
If a Premod or alternate electrotherapy treatment has been selected in combination with the Thermal treatment, the
electrotherapy treatment screen will remain in focus after pressing START. To move between the Electrotherapy and
Thermal display screens, press the Channels toggle key.