TeChnICal suMMary
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
129
4. PRESS and HOLD the START key for two full seconds to SAVE the new settings. A beep will sound signaling that
the new settings have been saved.
If the intensity was set before you pressed the START key, the treatment will begin upon pressing START. You may proceed
with delivering the treatment now, or you may stop the treatment.
Restore Factory Defaults
If you have saved your own defaults, but would like to return ALL the default settings to those that were set at the factory,
do the following:
1. Turn the machine off and wait five seconds.
2. Turn on the machine. Following initialization, DYNATRONICS appears in the Display Screen.
3. Press and hold the START key down until three beeps are sounded. The Factory Defaults have now been restored for
all modalities. You may now proceed with treatment setup.
4. To restore the factory defaults of only one modality, set the defaults for that modality to the settings referenced in the
operator’s manual. Press and hold the START key until a beep is heard.
Safety Features of the Dynatron Solaris PLUS
• Double redundancy protection on output amplifiers.
• Current sensing. If current reaches the current limit for the device, intensity is automatically reduced.
• All intensity levels are automatically set to zero at the end of treatment (ensures proper setting of intensity levels for
the next patient).
• Internal surge protection protects against line noise, machine switching operation and any other type of interference
that could cause patient discomfort.
• Soundhead temperature monitoring prevents the soundhead from becoming too hot, both to protect the soundhead
crystal from damage and to ensure patient comfort.