baTTery oPeraTIon
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
134
Battery Operation
All Solaris Plus Series devices are manufactured with battery capabilities allowing you to deliver treatments wherever power
may be unavailable or unreliable. To use the optional battery, do the following:
1. It is recommended that a battery be charged for 24 hours prior to operating the Solaris Plus device. DISCONNECT
the battery charging cable from the battery while it is in use for treatment.
2. Plug the battery adapter into the jack labeled “BAT-INPUT 12V-DC” on the back of the Dynatron Solaris Plus
console.
3. Set up and deliver treatments.
4. When battery power is reduced to approximately 11 volts, a low battery warning will be displayed in the lower
right-hand corner of the Treatment Display Screen “CAUTION: BATTERY LEVEL GETTING LOW!” The treatment
can continue however there will not be enough power to set up and deliver another treatment when the current
treatment has ended.
5. When the available battery power becomes too low to continue operating the device, the following message will
appear: “ERROR: BATTERY LEVEL TOO LOW FOR TREATMENT OPERATION. CHARGE IMMEDIATELY!”
Use ONLY a Dynatronics’ Approved Battery
Before purchasing or using an existing battery with a Solaris Plus device, contact Dynatronics or your Dynatronics
Representative to obtain specifications for a battery that may be safely used with a Solaris Plus device.
Only use a battery that CANNOT be recharged while it is in use. Disconnect the battery charger from the AC
power source before using the battery to supply power to this device.