eleCTroMagneTIC eMIssIons and IMMunITy
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
142
Electromagnetic
Emissions and Immunity
Tables 1 through 4 below list the Dynatron Solaris Plus declarations of electromagnetic emissions and immunity, and give
user guidance on the Dynatron Solaris Plus in an electromagnetic environment per IEC 60601-1-2 guidelines.
Table 1
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The dynatron solaris Plus (and accessories) is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the Dynatron Solaris Plus (and accessories) should assure that it
is used in such an environment.
emissions Test
Compliance
electromagnetic environment guidance
rf emissions
CIsPr 11
en55011
group 1
The dynatron solaris Plus (and accessories) uses rf en-
ergy only for its internal function. Therefore, its rf emis-
sions are very low and are not likely to cause any interfer-
ence in nearby electronic equipment
rf emissions
CIsPr 11
en55011
Class a
The dynatron solaris Plus (and accessories) is suitable
for use in all establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
harmonic emissions
IeC 61000-3-2
Class a
Voltage fluctuations/
flicker emissions
IeC 61000-3-3
Complies