fInal qualITy CheCK (qC) CheCKoff sheeT
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
148
Device Specific Procedure
QC, Assembly, Calibration, and Test
D719 Check Sheet
-
2 of 9-
DS0408A
–
REV 1
Refer to Test equipment operation manual and unit operating instructions/ manuals
REPAIR/RS
PROD
9
TRI-WAVE LIGHT MEZZANINE OUTPUT:
a)
Verify treatment selections work
b)
Verify light pad operation
c)
Verify light probe operation
Refer to Table for probe and pad treatment times
Wave length
Combination
DCP3
6 Joules
DLP3
10 Joules
IR / Red / Blue
1:00 min
8:42 min
10
Labels:
a)
Apply Serial Number Label
Earth/Enclosure/Patient Leakages(Enclosure/Patient: tester clip to surface of enclosure or Patient lead )
NORMAL
(no fault con neutral switch closed)
Earth Leakage
: Normal/Reversed Polarity, Earth switch depressed (<500 uA)
/
Enclosure Leakage
: Normal/Reversed Polarity, Chassis switch depressed(<100uA)
/
Patient Leakage
: Normal/Reversed Polarity, Chassis switch depressed(<100uA)
/
SINGLE FAULT
(fault condition/ neutral switch open)
Earth Leakage:
Normal/Reversed Polarity, Earth switch depressed(<1000uA)
/
Enclosure Leakage
: Normal/Reversed Polarity, Chassis switch depressed(<500uA)
/
Patient Leakage
: Normal/Reversed Polarity, Chassis switch depressed(<500uA)
/
ACCEPTANCE:
Initials:
Date:
Technician
QC Inspector