dynaTron solarIs® Plus PhysICal feaTures
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
15
16. Back Panel Jacks
a. Power 1/0 (ON/OFF) Switch. Located on the back of the unit this switch is labeled “1” and “0.” Set the switch
to “1” for ON; set the switch to “0” for OFF.
b. Battery. This jack may be used to supply power to the device using an optional battery pack. More information
about the optional battery operation is provided later in this manual.
c. SD Card Input. The SD Input provides a way for the Solaris Plus Devices to receive software updates quickly
and easily. Complete instructions for updating the devices using an SD card are found in the “Technical”
information section of the manual.
Current Limit
The Dynatron Solaris Plus devices continuously measure the actual current output during a treatment and limit the output
current to the level indicated in “Technical Information” in this manual. As the intensity of a treatment is increased the
current output is also increased.
When the maximum output current limit is reached, the device will immediately stop increasing the intensity and
automatically reduce the intensity a few increments to prevent the possibility of patient discomfort. Simultaneously, the
device will beep and one of the following CURRENT LIMIT WARNINGS will appear in the lower right-hand corner of the
Treatment Display Screen. Following is a list of CURRENT LIMIT WARNINGS that may occur.
Back Panel Jacks
c
b
a
NOTE: Patient Remote Stop. Adding the Remote Stop requires a custom order. The Patient Remote Stop Jack is
located below the Light Probe holder. The remote stop is controlled by the patient during unattended therapy, allowing
the patient to stop the treatment at any time. When the button on the remote stop cable is pressed, output for all Stim
modalities and pad treatments is stopped. During Combo treatments, both Sound and Stim outputs are stopped.
Remember to treat at the patient’s comfort level. It is not important to reach a given intensity level. It is only
important to set the treatment at a level that is comfortable to the patient. See “Electrotherapy Usage Cautions” in
this manual for suggested intensity limits.