dynaTron solarIs® Plus PhysICal feaTures
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
21
• With this method of re-hydration, after a couple of hours electrodes can regain up to 90 per cent of their original
adhesive quality.
2. NEVER use a self-adhesive electrode for more than 15 treatments (maximum).
3. NEVER USE STRAPS, WEIGHTS, or other devices to attach self-adhesive electrodes to the skin. If an electrode
has lost its adhesive quality, you can use one of the methods given above to re-hydrate the adhesive, or you should
discard the electrode. Using straps and weights with self-adhesive electrodes could have an unpredictable effect on
the electrodes and could cause injury.
4. NEVER use monitoring electrodes such as ECG, or EMG, nor ordinary TENS electrodes.
5. If you see the “No Patient Current” screen message, or if you observe poor conductivity indicators, check the
electrodes and lead wires for proper connection.