InTerferenTIal / PreModulaTed InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
27
Detailed Interferential / Premodulated Setup
1. IFC or PREMOD MODALITY keys
Press the IFC or PREMOD MODALITY keys to choose IFC (Interferential) or PREMOD (Premodulated). When
you choose IFC, two channels are automatically selected: 1-2 or 3-4. When choosing PREMOD, the first available
single channel is selected. Make sure the patient lead(s) is plugged into the correct jack(s) for the channel(s) selected.
The default settings for the modality are automatically selected. If you wish to use the default settings, increase the
intensity to the desired level, and press START.
Default Settings
Time:..........................................10 minutes
Frequency
Range: ...........High 80-150 Hz
Target
(IFC): .......................................... On
If you wish to change or customize the treatment settings, proceed through the following steps:
2. Customize TIME
The default treatment time is displayed at 10:00 min. Use the TIME arrow keys to increase or decrease the treatment time.
3. Customize FREQUENCY (optional)
Default HIGH and LOW Frequency Settings
• HIGH range is 80 to 150 Hz.
• LOW range is 0 to 10 Hz.
• ALTERNATING range alternates every 30 seconds between HIGH and LOW, beginning with LOW.
• CONSECUTIVE HIGH/LOW. During the first half of the treatment time the High frequency range is delivered.
During the second half of the treatment time the Low frequency range is delivered.
NOTE: Prior to increasing intensity, electrodes must be placed on the patient and the lead(s) attached to the device.
Plug the lead(s) into the channel(s) the device selects for this treatment. Consult published sources for electrode
placements, treatment settings, and treatment times. Make sure electrodes make good contact with the patient’s skin
over the entire surface area of the electrode. Improper electrode contact may result in patient injury.
NOTE: ALTERNATING and HIGH/LOW options may be selected after pressing START. However, Consecutive
and Fixed options must be selected before pressing START.