InTerferenTIal / PreModulaTed InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
29
SWEEP. The interferential current randomly sweeps the treatment area within the electrodes allowing the general
area to be bathed with Interferential current.
STATIC. The interferential current focuses only on the point where the current between electrodes intersects as it
follows the path of least resistance.
5. INTENSITY
Press the UP/DOWN arrow keys located next to the INTENSITY display on the right-hand side of the Treatment
Screen. When the INTENSITY is increased, current to the patient begins. NOTE: The CONDUCTANCE BAR
GRAPH will be operable at this time.
Before selecting the intensity setting for an individual patient, see “Electrotherapy Information and Usage Cautions”
in this manual for recommended intensity settings. Also see the section of this manual entitled “Contraindications,
Warnings, and Precautions” for specific precautions when treating any conditions contributing to loss of sensation,
or any time the patient cannot feel the electrical stimulation.
6. Press START
When you press start, the treatment timer in the TIME display window begins counting down and the treatment
proceeds. Remember to set the intensity before pressing START.
7. SAVE DEFAULTS
If the treatment you have just set up is a frequently used, you can save the treatment parameters as new defaults by
pressing and holding the START key until a beep sounds indicating the treatment parameters have been saved. The
next time you select the modality, these parameters will be selected automatically.
8. MODIFY SETTINGS
Treatment settings can be modified while the treatment is in progress except for a CONSECUTIVE HIGH/LOW
treatment. If the TIME setting on a CONSECUTIVE HIGH/LOW treatment is altered, the treatment will be aborted
and default to an ALTERNATING HIGH/LOW treatment.
• FREQUENCY RANGE. Use the Range Toggle key to select a different frequency option (High, Low, Alternating
High/Low, Consecutive High/Low, or Fixed). Fixed Frequency and Consecutive High/Low cannot be selected
after treatment is started.
• TARGET/SWEEP/STATIC. Use the Target Toggle key to select Target, Sweep, or Static (for IFC only).
• TIME. Use the Time Arrow keys to increase or decrease the treatment time.
• INTENSITY. Use the Intensity Arrow keys to increase or decrease the intensity.
• TARGET. Relocate the treatment site by touching the TARGET PAD at any time during the treatment when
TARGET has been selected.