InTerferenTIal / PreModulaTed InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
30
9. STOP.
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. This stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. The output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).