hIgh VolT InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
46
allowing for flexible time for the delivery of current to each treatment site. When you press the button on the probe
to STOP the current, the device beeps twice and the timer resets to zero.
4. PULSE RATE RANGE.
Select a PULSE RATE RANGE (High 80-120 Hz or Low 1-10 Hz) or set a single pulse rate by pressing the MODE
toggle key and selecting RATES from 1-120 Hz. Use the arrow keys next to RATE-START and RATE-STOP to set
the Pulse Rate Range. These ranges may be changed for a single treatment if desired, or press and hold the START
key to set new default settings to be applied to all future treatment setups. The PULSE RATE RANGE must be set
before pressing START.
5. CONTRACTION (ON) and REST (OFF) TIMES
FUNCTION KEY: To access the CONTRACTION AND REST settings, press the FUNCTION key.
Press the CONT/REST toggle key one or more times to select contraction/rest (Duty) cycle times. Available options
include 10/10, 10/30, 10/50, CONT (Continuous). and Custom. The setting of 10/30, for example, means a 10-second
contraction time followed by a 30-second rest time.
CUSTOM (Duty) CYCLE TIME SELECTIONS
• Press the FUNCTION key located on right-side of the device faceplate.
• Using the CONT/REST toggle key, select CUSTOM in the CONT/REST window.
• Using the MODE toggle key, select CUST CR in the MODE window. Selections will be illuminated GREEN.
• Custom Contraction and Custom Rest cycle times may now be set by using the Up and Down arrow keys
located next to Custom Contraction and Custom Rest windows. Available ranges for Contraction (ON) times
are 1-120 seconds, for the REST (OFF) times 1 to 300 seconds. Remember, the REST time cannot be less than the
Contraction time. Pressing and holding the START key until the beep is heard will save the current Contraction/
Rest settings as the default.
6. Choose the RAMP setting.
FUNCTION KEY: To access the RAMP settings, press the FUNCTION key.
Press the FUNCTION key on the right side of the face plate. Press the RAMP toggle key to select RAMP time.
Selections include: 0.5, 1.0, 1.5, and 2.0. Ramp time is applied to both the start and end of the contraction. Ramp
time is in addition to the contraction time itself. A ramp setting is not applied to the Continuous duty cycle except
when using a probe. With a probe treatment set to a Continuous duty cycle, a ramp time of 3 seconds is automatically
applied.
7. INTENSITY.
Using the Up and Down arrow keys to the right-side of the INTENSITY display, set the Intensity. Remember, when
you increase intensity, current to the patient begins. Therefore, START should be pressed immediately after setting
the intensity to begin the treatment timer.