MICroCurrenT ModalITy InforMaTIon
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
57
Probe. If Probe treatment is selected, the following defaults are automatically selected.
Note: Bipolar is not available in Probe mode.
Polarity ...................................................................................Negative Unipolar
Frequency ..................................................................................................... 4 Hz
Intensity .....................................................................................................300µA
Time ................................................Determined by Practitioner - Counts Up
Audible Tone ...................................................................................................On
Conductance ...................................................................................................On
Available Ranges.
Frequency ........................................................................................ Between 0.1 and 200 Hz (positve or negative)
Biploar 0.0-5.0 Hz
Maximum Voltage: .......................................................................................................................60 V (open circuit)
Intensity: ..........................Between 10 and 990 microamps open load, constant current, increments of 10 µA.
Maximum Load Resistance ........................................................................................................................ 500 Ohms
The ranges of .1 to 200 Hz frequency and between 10 and 990 micro amperes intensity are available for setting up both
electrodes and probes treatments. For guidance in selecting the appropriate frequency and intensity for a given probes or
electrodes treatment, consult published medical literature.