Dynatron Tri-Wave
™
Light Quick Treatment Setup
1. Remove all gels and lotions from skin surface.
2. While the Dynatron Solaris Plus controlling console is turned OFF, plug the Dynatron Tri-Wave Light
probe and/or Dynatron Tri-Wave Light pads into the corresponding output jacks located on the side(s)
of the console.
3. Placing protective eyewear on both the patient and the practitioner is required.
4. Turn ON the Dynatron Solaris Plus controlling console.
5. LT PROBE and LT PADS. Press Probe or Light Pad on the Solaris Plus console. When using the 707, press
the toggle key below the channels window to select the desired Probe or Pads to activate.
6. Select WAVELENGTHS. Using the Wavelength toggle keys, select the desired wavelengths(Infrared,
Blue, Red) by toggling the wavelength ON or OFF. The selection will be illuminated in GREEN when the
wavelength is active. Note: Wavelengths cannot be changed during a treatment.
7. TIME/DOSAGE. Using the Up/Down arrow keys located next to the TIME and DOSAGE displays,
Dosage/Time may be adjusted. Dosage and Time are calculated according to the wavelengths selected. If
changes are made to Dosage, Time, will automatically be recalculated. If changes are made to Time, Dosage
will automatically be recalculated.
8. START. Press START on the Dynatron Solaris Plus control console.
Probe. After pressing START on the console, press the 1/0 (ON/OFF) key located on the Probe handle to
begin treatment.
Pads. The Pads treatment begins immediately when the console START key is pressed.
9. PAUSE.
Probe. Press the 1/0 (ON/OFF) key located on the Probe handle to pause a Probe treatment or press the
PAUSE key located on the console. To continue the treatment, press the 1/0 (ON/OFF) key on the Probe
handle or the PAUSE key on the console again.
Pads. With the desired set of Pads as the focus treatment, press the PAUSE key located on the Solaris Plus
console. To continue the treatment, press the PAUSE key again.
10. STOP
Probe. Press the 1/0 (ON/OFF) key on the Probe to pause or stop a probe treatment or press the FUNCTION/
STOP keys on the console simultaneously.
Pads. Press the FUNCTION/STOP keys on the console to immediately STOP a Pads treatment, reduce the
treatment time to zero or allow the treatment to automatically time out.
NOTE: Pressing only the STOP key on the Dynatron Solaris Plus console will terminate all treatments in progress.