dynaTron TrI-WaVe™ oPeraTIng InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
66
treatment, press the TREATMENT MODALITY key for the desired treatment. When a treatment is in focus, the
selected treatment will be highlighted in GREEN text within the Channels window and the treatment name will
appear in the center of the Treatment Display Screen.
5. Select WAVELENGTHS (Total Energy Delivered)
The user has the capability of selecting a specific wavelength or blending various wavelengths into a single treatment.
Wavelengths are selected by using the WAVELENGTH TOGGLE keys to turn the Wavelengths “ON” and/or “OFF.”
The default wavelength for the Probe and Pads is Infrared (850nm) and Red (624nm).
To select a wavelength other than the Infrared (850nm) default, press the key under the light therapy option you wish
to activate: Red (624nm), Blue (464nm), or Infrared (850nm). Pressing the key under each option will cause the ON
text to become illuminated. Pressing the key again will cause the OFF text to illuminate. Selections may be activated
individually or in any combination.
Dosage (J/cm
2
) is calculated as a combination of Time and Power, and is automatically calculated once the Wavelength
selections have been made. Dosage can manually be altered by using the Up/Down arrow keys located on the
right-side of the Dosage display. When Dosage is altered, treatment time is automatically adjusted.
Dynatron Tri-Wave
™
Light Wavelength Settings
The following tables list wavelength selections or combination of selections with their respective dosage and treatment times.
TRI-WAVE LIGHT PROBE DEFAULTS
WaVelengThs
defaulT dosage
PoWer ouTPuT mW
TIMe
Infrared (Ir) – (850nm)
6 j/cm
2
1000 mW
35 sec.
blue – (464nm)
6 j/cm
2
500 mW
1 min. 12 sec.
red – (624nm)
6 j/cm
2
500 mW
1 min. 12 sec.
Infrared and red
6 j/cm
2
800 mW
44 sec.
blue and Infrared
6 j/cm
2
800 mW
48 sec.
Infrared, blue and red
6 j/cm
2
650 mW
1 min.
blue and red
6 j/cm
2
525 mW
1 min. 11 sec.
NOTE: Wavelength selections cannot be changed during a treatment.