33
Rev. 0.6
info@e-breathe.de • www.e-breathe.de
e-breathe e-Flow
3.1 Blower Unit
3. Technical Specifications
3.2 Battery
3.3 Charger
Blower operation mode:
Approvals:
e-breathe e-Flow Hood System EN 12941 (TH2 / TH3)
e-breathe e-Flow Full Face Mask System EN 12942 (TM2/TM3)
air flow rate
(automatic
readjustment):
Adjustable on the device in three stages:
160 - 180 - 200 l/min Hood-System
120 - 140 - 160 l/min Full Face Mask-System
Airflow warning:
Battery warning:
< 160 l/min Hood-System
< 120 l/min Full Face Mask-System
< 25-15 min remaining operating time or < 5% battery capacity
Alarm system:
Optical alarm (indication at the colour display with error code)
Audible alarm (≥ 75 dB )
Vibration alarm
Nominal operating time:
min. 4 hours - approx. 8 hours
(depending on the adjusted air flow and the connected filters and head piece)
Temperature range:
-10°C to +40°C <70% relative humidity
Storage temperature:
0°C to +30°C <70% relative humidity
Battery type:
Lithium-Ion
Charging time:
< 2,5 hours
Temperature range:
-10°C to +40°C <70% relative humidity
Storage temperature:
0°C to +30°C <70% relative humidity
Charging temperature:
0°C bis +30°C
Nominal voltage:
14,4 V
Nominal capacity:
3,4 Ah
Power output:
49 Wh
Shelf life:
Fully recharge every six months or at least once a year.
Charging cycles:
300-500
Input voltage:
18 V
Input stream
2,22 A
Protection class:
IP 30
Storage temperature:
0°C bis +30°C <70% relative humidity
The battery fulfills the following regulations:
EMC Directive (2004/108/EC), RoHS Directive (2002/95/EC), Battery Recycling Directive (2006/66/EC), NSR Directive (2006/95/EC)
The battery is approved according to the following standards:
CE [EN55022:2006 (ITE Class B) & EN55024:1998 (ITE)], FCC Part 15 Class B