4.2.4 Battery Assembly / Replacement
4.2.4.1 Battery Disassembly
The battery locking latch is located on the right side of the blower when it stands in front of you. Press the latch to
release the battery from the unit. Make sure that the battery does not fall down. Remove the battery.
See Figure 9.
4.2.4.2 Battery Assembly
Lead the battery diagonally with the contacts to the blower connection. Make sure that the battery is in the battery tray and
that the contacts of the battery and the connections on the blower are in contact. Insert the battery so that the battery and
the battery locking button are audibly and noticeably connected. As a precaution, check that the battery is firmly inserted
in the battery compartment.
See Figure 10.
Important: Damage to the battery due to short circuit possible!
The battery may be damaged by a short-circuit. The battery must always be placed in such a way that the con-
tacts do not come into contact with metal or other materials that could cause a short-circuit.
Never attempt to disassemble the battery.
4.2.5 Battery Charging
Important: The battery can only be charged in the corresponding charging station from e-breathe. In addition,
the battery can only be charged if it is intact and not damaged.
If you are not using the charger, disconnect it from the power supply. Protect the charger from sunlight, direct
heat radiation, dust, dirt and moisture.
4.2.5.1 The correct mains voltage of the power supply must be checked. The battery voltage of the power supply unit
must match the mains voltage.
4.2.5.2 If the voltage is correct, connect the power supply unit to the power supply.
4.2.5.3 Insert the battery into the charging station. Make sure that the contacts of the battery meet the plugs of the
charging station. Be careful while inserting the battery so that neither the battery nor the charging station are da-
maged.
4.2.5.4 Wait for the charging process to proceed. As long as the green LED on the charging station flashes, the battery
is in charging mode. As soon as the green LED stops flashing, the battery is fully charged.
After charging, remove the battery and disconnect the mains adapter from the power supply when it is no longer in
use.
4.2.5.5 If you additionally want to check the capacity of the battery, you can mount the battery in the blower, switch
on the device and read the capacity of the battery in the display.
Batteries that are stored should be recharged first.
If the batteries are stored for more than 6 months, they should be fully recharged in the meantime.
Important: The charging station (incl. power supply) must also be completely intact before it can be operated. If
the power supply unit is damaged, it cannot be used under any circumstances! The charging station can only be
used in closed rooms. During charging, the power supply unit should not be covered, located near heat radiation
or exposed to direct sunlight.