41
Rev. 0.6
info@e-breathe.de • www.e-breathe.de
e-breathe e-Flow
160 l/min
100%
10:27
performing self-check
5.0 e-Flow Functional Description
By pressing the power button/menu button (round button) the blower is switched on. By pressing the power button
for 2 seconds, the blower is switched off again.
Upon switching on, an acoustic signal sounds briefly, the vibration motor responds briefly and the display lights up.
5.1 System Test
Afterwards the blower performs a self-check. In this phase, the start logo
is displayed. This process takes about 5 seconds.
If an error occurs during the Self-Check phase, an alarm is shown on the
display with the corresponding error code for 5 seconds. The unit then
switches itself off automatically to prevent the blower from starting up
without correcting the error.
If no error occurs during the Self-Check phase, the display automatically
switches to the main menu.
The main menu displays the following information:
- default air flow mode
- Required number of filters
- battery capacity
- remaining battery runtime
- filter capacity
- Adjusted air flow stage 1- 3
5.2 Main Menu (1)
5.2.1 Default air flow mode
Hood Mode 160-180-200 l/min
Full Face Mask 120-140-160 l/min
The blower has 2 modes which can only be changed via a PC software application.
For more information, please contact www.e-breathe.de.
5.2.2 Number of filters:
The number of required filter. One filter, per filter shown in the status line, must be connected to the
blower.
5.2.3
Battery capacity and remaining battery runtime
Battery capacity and remaining battery runtime.
For the first 3 minutes, the battery capacity is displayed in percent. Afterwards the remaining battery runtime in h/min
is displayed. After that the battery capacity is only displayed by its colour and the number of bars.
Display Symbol
Meaning
- 4 green bars = > 75%
- 3 green bars = > 50%
- 2 yellow bars = > 25%
- 1 yellow bar = > 15%
- 1 red bar = < 15%