45
Rev. 0.6
info@e-breathe.de • www.e-breathe.de
e-breathe e-Flow
Meaning
Cause of error
Action
Display Symbols
Alarm Battery capacity
exhausted
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:22 / Err:23
The remaining battery run-
time is exhausted < 20-15
min and/or the battery ca-
pacity is < 5%.
At 0% / 0 min battery capa-
city the blower will switch
off for self-protection.
On switch-on:
Battery must be charged.
During operation:
Work must be interrupted
and the battery must be
charged or replaced with a
fully charged battery.
Alarm Filter capacity
exhausted
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:40
The filter capacity is < 10
%, the minimum volume
flow cannot be maintained
in the foreseeable future
due to further saturation
of the filter.
After switching on:
Filters must be changed.
During operation:
Work must be interrupted
and the filters must be re-
placed with new ones.
Alarm Motor
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:10 Motor load or mo-
tor speed is too high, the
minimum air flow can no
longer be achieved.
Err:11 Motor is blocked. No
volume flow can be gene-
rated.
Err:13 Incorrect volume
flow output. Unusual pow-
er consumption.
Work must be interrupted
and the filters must be re-
placed with new ones.
Make sure that there are no
obstacles in the air outlet or
duct and remove them.
Restart the blower and test
the volume flow with the
test tube. Have blower tes-
ted by e-breathe service.
Alarm Service Date
• When switching on
• Duration 10 sec.
repeated
• Acoustic alarm tone
• Optical alarm
• Vibration alarm
Can‘t be switched off
Err:50 The next scheduled
service date has been ex-
ceeded. The alarm is trig-
gered for 10 seconds each
time the unit is switched
on until the service has
been carried out.
The device must be main-
tained by the manufacturer
or an authorized service
partner.
Alarm System Error
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can‘t be switched off
Err:70 Communication
problems with the air flow
sensor.
Have blower tested by
e-breathe service.
00:35
Err:22
Err:40
Err:10
05.11.2019
Err:50
Error
Err:70