e-breathe e-Flow
info@e-breathe.de • www.e-breathe.de
e-breathe e-Flow
5.9 Warnings and Alarms
The blower is equipped with the following warning devices to safely protect and inform the user.
- Audible alarm tone
- Optical alarm through flashing of the display
- Optical alarm through indication of the error on the display
- Mechanical alarm/vibration alarm
Since, depending on the activity or working area, you may not be able to detect one of the warning alarms, all war-
ning alarms are always displayed simultaneously in the event of an alarm.
5.9.1 Warning Devices
5.9.2 Warnings
A warning always serves as prior information for the user. A warning tells the user that there will be an alarm shortly
afterwards, leading to an interruption of work.
In the event of a warning, the work does not have to be interrupted.
A warning is signalled when the warning devices „audible alarm tone“ and „visual alarm“ are activated. The duration
of a warning is 10 seconds, afterwards it stops. A warning can be prematurely confirmed by briefly pressing the menu
button.
In the case of a warning, a warning triangle is shown on the left-hand side of the display and the reason for the war-
ning is shown on the right-hand side.
Meaning
Cause of error
Action
Display Symbols
Warning Battery Capacity
low
• In operation
• Duration 10 sec. - once
• Acoustic alarm
• Optical alarm
Can be switched off
Remaining battery runti-
me is low < 40-30 min and/
or battery capacity is < 8%.
Work must be interrupted
shortly and the battery
must be charged or repla-
ced by a fully charged bat-
tery.
Warning Service Date
• In operation
• Duration 5 sec. - once
• Acoustic alarm
• Optical alarm
Can‘t be switched off
The user is reminded of
the next due service date
at preset intervals.
A service appointment with
the manufacturer or an
authorized service partner
should be made shortly.
00:35
05.11.2019
5.9.3 Alarms
In the event of an alarm, work must be interrupted, and the dangerous area must be left.
An alarm is signalled when the warning devices „audible alarm tone“, „visual alarm“ and the vibration alarm are activa-
ted. An alarm always sounds continuously or as long as the alarm reason/error reason still exists. Some alarms can be
confirmed by briefly pressing the menu button. The alarm will stop for xx seconds and start again if the alarm cause
still exists.
Alarms or errors that occur during the self check phase are displayed for 5 seconds. The blower automatically switches
off after this period, so that operation is not possible without eliminating the fault.
In the event of an alarm, a skull and crossbones with the corresponding error code is shown on the left-hand side
of the display and the reason for the alarm on the right-hand side. If several alarms occur at the same time, they are
displayed alternately.
Meaning
Cause of error
Action
Display Symbols
Alarm Battery capacity
exhausted
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:22 / Err:23
The remaining battery run-
time is exhausted < 20-15
min and/or the battery ca-
pacity is < 5%.
At 0% / 0 min battery capa-
city the blower will switch
off for self-protection.
On switch-on:
Battery must be charged.
During operation:
Work must be interrupted
and the battery must be
charged or replaced with a
fully charged battery.
Alarm Filter capacity
exhausted
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:40
The filter capacity is < 10
%, the minimum volume
flow cannot be maintained
in the foreseeable future
due to further saturation
of the filter.
After switching on:
Filters must be changed.
During operation:
Work must be interrupted
and the filters must be re-
placed with new ones.
Alarm Motor
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can be switched off
Err:10 Motor load or motor
speed is too high, the mini-
mum airflow can no longer
be achieved.
Err:11 Motor is blocked. No
volume flow can be gene-
rated.
Err:13 Incorrect volume
flow output. Unusual pow-
er consumption.
Work must be interrupted
and the filters must be re-
placed with new ones.
Make sure that there are no
obstacles in the air outlet or
duct and remove them.
Restart the blower and test
the volume flow with the
test tube. Have blower tes-
ted by e-breathe service.
Alarm Service Date
• When switching on
• Duration 10 sec.
repeated
• Acoustic alarm tone
• Optical alarm
• Vibration alarm
Can‘t be switched off
Err:50 The next scheduled
service date has been ex-
ceeded. The alarm is trig-
gered for 10 seconds each
time the unit is switched
on until the service has
been carried out.
The device must be main-
tained by the manufacturer
or an authorized service
partner.
Alarm System Error
• On switch-on 5 sec.
• Permanent during
operation
• Acoustic alarm
• Optical alarm
• Vibration alarm
Can‘t be switched off
Err:70 Communication
problems with the airflow
sensor.
Have blower tested by
e-breathe service.
00:35
Err:22
Err:40
Err:10
05.11.2019
Err:50
Error
Err:70