8
User Manual for EE08 Humidity / Temperature Probe
7 Accessories / Spare Parts
M12 connection cable for type E11, length 1.5 m
(5 ft)
HA010322
M12 connection cable for type E11, length 3 m
(10 ft)
HA010323
M12 connection cable for type E11, length 5 m
(16.4 ft)
HA010324
M12 connection cable for type E11, length 10 m
(32.8 ft)
HA010325
Radiation shield for Type E8
HA010502
Radiation shield for Type E11
HA010506
Wall mounting clip Ø12 mm
HA010211
Protection cap for Ø12 mm probe
HA010783
M12 female socket with wires
HA010703
M12 female cable connector assembly possible
HA010704
Metal grid filter
HA010113
Cconfiguration cable
HA011005
EE-PCS
free download at www.epluse.com/EE08
8 Technical Data
Measurands
Relative Humidity
Measuring range
0...100 % RH
Accuracy
at 23 °C
(73 °F)
for RH ≤ 90 %
±2 % RH
and nominal voltage
1)
f
or RH > 90 %
±3 % RH
Temperature dependence, typ.
0.03 % RH/°C
(0.02 % RH/°F)
Temperature
Measuring range
-40...80 °C
(-40...176 °F)
Accuracy
at nominal voltage
1)
Outputs
Analogue
0 - 1 V / 0 - 2.5 V / 0 - 5 V / 0 - 10 V
-0.2mA < I
L
< 0.2 mA
Digital interface
E2 interface
2)
General
Supply voltage
for output 0 - 1 V / 0 - 2.5 V
V1: 4.5 - 15 V DC V2: 7 - 30 V DC
for output 0 - 5 V
V2: 7 - 30 V DC
for output 0 - 10 V
V2: 12 - 30 V DC
Current consumption, typ.
< 1.3 mA
Electrical connection
M12x1, 8 poles
Cable PVC 8 x 0.14 mm² (M1 models)
Cable PVC 10 x 0.14 mm² (M6 models)
Filter
Metal grid
Protection rating
IP65
Enclosure material
Polycarbonate
Electromagnetic compatibility
EN 61326-1
EN 61326-2-3
(Industrial Environment)
Operating and storage conditions
-40...80 °C
(-40...176 °F)
0...100 % RH (operation)
0...95 % RH non-condensing (storage)
Adjustment
With EE-PCS (Product Configuration Software, free download)
and configuration adapter
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated
in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement); nominal voltage V1 = 12 V DC, V2 = 24 V DC
2) For further support literature refer to www.epluse.com/ee08.