4

LED INDICATION

Factory Setup: 

The display shows the two parameters selected for output 1 and output 2 (according to ordering 

code). For digital output versions the display shows RH and T.

User Setup: 

The user can change the display layout to 1, 2 or 3 lines and select the  

parameters to be displayed by using EE-PCS Product Configuration Software (free download from 

www.epluse.com/configurator) and the optional USB configuration adapter (not included in the 

scope of supply).

Important:  

The EE210-M1xA6 (4...20 mA, 2-wire version) with display operates correctly only if 

both outputs are connected. 

DISPLAY

SELECTION OUTPUT SIGNAL U/I

The factory setup of the output signal and scaling corresponds to the type number as ordered.

The output signal (voltage or current 3-wire) can be selected with the DIP switch on the main electronics board (see picture PCB 

EE210-M1xA2/3/5). This does not impact on the scaling of the outputs, which can be changed using USB configuration adapter and 

the EE-PCS. 

Examples

Factory setup: voltage output 0 - 5 V or 0 - 10 V corresponds to 0…100%RH. 

After switching from U to I: current output 0…20 mA corresponds to 0…100%RH.  

A change of the current output range for instance to 4 - 20 mA (3-wire) can be made subsequently with the USB configuration adapter 

and EE-PCS.

Factory setup: current output 0 - 20 mA corresponds to -10…50 °C.

After switching from I to U: voltage output 0 - 10 V corresponds to -10…50 °C.  

A change of the voltage output range for instance to 0 - 5 V can be made subsequently with the USB configuration adapter and 

EE-PCS.

Green LED

Information during normal operation

Blue LED

 

Information during setup with the optional USB  

configuration adapter HA011066 and the EE-PCS  

Product Configuration Software

on

= everything OK

on

= USB stick connected to PC, no communication 

with EE-PCS

flashing = the main board does not recognize the  

measurement electronics inside the RH probe

flashing = communication in progress with EE-PCS

off

= no power supply or main board failure

off

= USB stick not connected to PC

HARDWARE BUS TERMINATION

If required, the bus termination shall be realized with 120 Ohm resistor, jumper on the board.

•  Jumper mounted 

= bus terminated

•  Jumper not mounted  = bus not terminated

ADDRESS SWITCH

Address Switch

Address Switch

Address setting via EE-PCS Product Configuration Software

All DIP switches at position 0 → address has to be set via

 

Product Configuration Software

Factory setting: 242 (permitted values: 1…247).

Example: Slave address is set via configuration software.

Address setting via DIP switch

Setting the DIP switches to any other address than 0, overrules the  

slave address set via configuration software (permitted values: 1…247).

Example: Slave address set to 11 (= 0000 1011 binary).

0 0 0 0 0 0 0 0

0
1

1 1 0 1 0 0 0 0

0
1

DIGITAL SETTINGS

Summary of Contents for EE210

Page 1: ...or and mainly the sensing head shall not be exposed to extreme mechanical stress The sensor must be operated with the filter cap on at all times Do not touch the sensing element inside the sensing hea...

Page 2: ...10501 for Type T13 needs to be ordered separately Type T13 5 20 20 1 5 122 44 105 174 44 94 4 1 6 9 1 7 3 7 4 8 0 2 0 8 0 1 0 8 1 7 80 37 2 80 55 32 12 1 5 3 5 3 5 2 2 1 3 0 5 EE210 with cable gland U...

Page 3: ...t The replacement probe shall be installed as described above Important Make sure that the cable glands are closed tightly for both PE210 probe cable and for the power supply and outputs cable This is...

Page 4: ...y setup current output 0 20 mA corresponds to 10 50 C After switching from I to U voltage output 0 10 V corresponds to 10 50 C A change of the voltage output range for instance to 0 5 V can be made su...

Page 5: ...0 0x27 specific enthalpy kJ kg BTU lb 30042 0x29 frost point temperature C F INTEGER 16 bit read register Function code Register number1 Dec Register address2 Hex Parameter name Unit3 Scale4 30301 0x1...

Page 6: ...2 Register 2 Lo Byte 3 Register 1 Hi Byte 4 Register 1 Lo 41 C2 29 00 24 270019 TECHNICAL DATA Modification rights reserved Measured Values Relative Humidity RH Working range 0 100 RH RH accuracy1 inc...

Page 7: ...y max 40 mA with display max 40 mA 3 wire DC supply typ 33 mA with display max 44 mA AC supply typ 65 mArms with display max 84 mArms Digital interface DC supply typ 5 mA with display max 20 mA AC sup...

Page 8: ...nik Calibration and or adjustment can be performed in the E E Elektronik calibration laboratory For information on the E E capabilities in ISO or accredited calibration please see www eplusecal com Ca...

Page 9: ...to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates...

Reviews: