19
Electrical connections
Connection configuration of bottom part of the housing with plug connections
8...35V DC / 12...30V AC (option C03/C07)
plug for supply and
analogue output
(front view)
1
2
3
4
5
Description:
Connection assignment:
V+
5
GND
4
GND
3
OUT1
2
OUT2
1
Connection configuration of bottom part of the housing with integrated power supply
100...240V AC (option V01)
1
2
3
4
5
Euro-Standard
Euro-Standard
plug for
analogue outputs
(front view)
plug for 100-240V
metal housing
(front view)
plug for 100-240V
polycarbonate housing
(front view)
Description:
Connection assignment:
GND
3
OUT1
2
OUT2
1
Description:
Connection assignment:
grounding (PE)
1
phase (L1)
2
neutral wire (N)
3
Description:
Connection assignment:
phase (L1)
1
neutral wire (N)
3
External diameter of supply cable: 10-12mm
(0.39-0.47”)
Maximum wire cross section: 1,5mm² (AWG 16)
The protection of the supply cable against excess current and short-circuit shall be in
accordance with national and local codes.
Bottom and centrepiece of the housing shall be grounded!
!