13
Operating manual EE23 humidity / temperature sensor
Pin assignment for option E4 (with supply 8...35V DC / 12...30V AC)
plug for supply and
analogue output
(front view)
1
2
3
4
5
Description:
Pins:
V+
5
GND
4
GND
3
OUT1
2
OUT2
1
Pin assignment for option AM3 (integrated supply unit for 100...240V AC)
plug for
analogue outputs
(front view)
1
2
3
4
5
Description:
Pins:
GND
3
OUT1
2
OUT1
1
plug for 100-240 V
metal enclosure
(front view)
Description:
Pins:
grounding (PE)
1
phase (L1)
2
neutral wire (N)
3
plug for 100-240 V
polycarbonate enclosure
(front view)
Description:
Pins:
phase (L1)
1
neutral wire (N)
3
External diameter of supply cable: 10-12mm (0.39-0.47”)
Maximum wire cross section for connecting cable: 1.5mm² (AWG 16)
The protection of the supply cable against excess current and short-circuit must be designated
to a wire cross section of 0.8mm² (AWG 18) (6A fuse).
The back cover and the middle part of the metal enclosure must be grounded during operation!