Example (b): „Mapping an EE07 sensing probe with serial number 0909500001055D“
Assigned to:
Select “Probe”.
Serial Number:
Enter the serial number of the sensing probe. The serial number can be copied and
paste from the “Transmitter Details” of the transmitter list see chapter
12.2.1 Transmitter List with ‘Ctrl’+’C’ and ‘Ctrl’+’V’.
Transmitter Port:
Enter „255“.
2. Setup of ranges and values:
Measurand and Unit: Selected desired measuring value (T, RH …).
Physical Range:
Enter desired range and engineering unit of the measured value (e.g. 0 ...100 °C).
Output Type:
Indicates the physical output quantity (factory setting according to order code).
Custom Range:
Analogue output value, to represent the “Physical Range” as indicated above.
Failsafe Value:
Default analogue output value if a transmission failure / alarm is present.
12.4 Modbus Map
Via the link „Add new Modbus Register“, new registers / variables can be created.
Register Number:
Is incremented automatically, but can be changed any time.
Assigned to:
Here you select whether a sensor or a transmitter will be mapped to the register.
(Details see 12.3. Outputs → Mapping the selected measurement signal)
Serial Number:
Type in the serial number of the probe or transmitter.
(can be copied from the transmitter list (see section 12.2))
Transmitter Port:
Type in transmitter port (see section 12.2.2)
Measurand / Unit:
Adjust desired measurement (T, RH ,..).
Failsafe Value:
Output value, which should be issued for a (transmission) error/alarm.
Data Type:
Select data type (Float, Integer,..)
Factor:
Possible multiplication factor for the register value
(Reg.Value = Current Value * Factor)
Offset:
Possible offset for the register value (Reg.Value = Current Value * Offset)
Create Register:
The register will be created with the selcted configuration.
25
Operating manual Series EE240 Wireless sensor