background image

13

1.

Insert the sensor probe into the reference temperature 1 (low

calibration point ) and stabilise for at least 10 minutes.

2. BUTTON S1: 

Pressing the button for 5 seconds starts the

procedure for the calibration mode temperature. The calibration
mode is indicated by the lit LED "D1" on the circuit board. 

3. BUTTON S2: 

Pressing the button for 5 seconds starts the

procedure for the low calibration point. The calibration mode is
indicated by the symbol "CALIB LOW" on the optional LCD
display.

4. BUTTON S1 (up) 

and

S2 (down): 

Pressing one of the two

buttons will adjust the measuring value in steps of 0.1 degC up
or down to the reference value. The actual measuring value is
indicated on the display or can be measured with the analogue
output. As soon as the measured value is changed, "D1" is flas-
hing.

5. BUTTON S1 (store): 

Pressing the button for 5 seconds 

stores the calibration value and the procedure is ended. LED
"D2" flashes to indicate exiting of the calibration mode and the
symbol "CALIB LOW" will disappear from the optional LCD
display.

BUTTON S2 (cancel)

: Pressing the button for 5 seconds the

calibration procedure will be ended without storing the 
calibration values. LED "D2" flashes to indicate exiting of the
calibration mode and the symbol "CALIB LOW" will disappear
from the optional LCD display.

low calibration point:

6.

Insert the sensor probe into the reference temperature 2

(high calibration point) and stabilise for at least 10 minutes.

7. BUTTON S1: 

Pressing the button for  5 seconds starts the

procedure for the calibration mode temperature. The calibra-
tion mode is indicated by the lit LED "D1" on the circuit board. 

8. BUTTON S1: 

Pressing the button for 5 seconds starts the

procedure for the high calibration point. The calibration mode
is indicated by the symbol "CALIB HIGH" on the optional LCD
display.

9. BUTTON S1 (up) 

and

S2 (down): 

Pressing one of the two

buttons will adjust the measuring value in steps of 0.1 degC
up or down to the reference value. The actual measuring value
is indicated on the display or can be measured with the analo-
gue output. As soon as the measured value is changed, "D1"
is flashing.

10. BUTTON S1 (store): 

Pressing the button for 5 seconds 

stores the calibration value and the procedure is ended. LED
"D2" flashes to indicate exiting of the calibration mode and the
symbol "CALIB HIGH" will disappear from the optional LCD
display.

BUTTON S2 (cancel)

: Pressing the button for 5 seconds the

calibration procedure will be ended without storing the cali-
bration values. LED "D2" flashes to indicate exiting of the cali-
bration mode and the symbol "CALIB HIGH" will disappear
from the optional LCD display.

2-point temperature calibration procedure on the circuit board!

high calibration point:

S2

S2

S1

S1

S1

S2

D1

red

D1

red

D1

red

D2

D2

“CALIB LOW”

S1

S1

S2

S1

S1

S2

D1

red

D2

D2

“CALIB HIGH”

Hardware

Summary of Contents for EE29 Series

Page 1: ...UAL Hardware and Software Hardware and Software Series EE29 EE31 Series EE29 EE31 HUMIDITY TEMPERATURE HUMIDITY TEMPERATURE TRANSMITTER TRANSMITTER BA_EE29_EE31_05_e technical data are subject to chan...

Page 2: ...used in accordance with the installation manual may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If t...

Page 3: ...tion 11 7 2 2 point temperature calibration 12 7 3 1 point humidity calibration 14 7 4 1 point temperature calibration 15 7 5 Resetting to factory calibration 16 8 MAINTENANCE 17 8 1 Sensor replacemen...

Page 4: ...ilter cap as the sensor element could be damaged The sensor is an Electro Static Discharge sensitive component ESD When touching the sensor element ESD protective measures should be followed Grip sens...

Page 5: ...Various options extend the large number of applications The construction of the transmitter enables field calibration and local loop calibration in a simple way 5 3 INSTALLATION Mounting of the housi...

Page 6: ...lly or vertically pointing downwards When possible a drip sheet should be fitted for each mounting i stainless steel mounting flange bow for draining water of condensation 13 vertical horizontal mount...

Page 7: ...Rel 2 4 ELECTRICAL CONNECTIONS 1 2 3 4 5 1 2 3 4 5 Description Connection assignment V 5 GND 4 GND 3 OUT1 2 OUT2 1 Description Connection assignment GND Ser 5 Rxd B 3 Txd A 1 not assigned 2 4 Euro St...

Page 8: ...e removed 3 Fitting of the network chip For refitting to RS485 an IC must be used available as an option The notch on the chip must match the receiver slot 4 Display These pinboards are determined for...

Page 9: ...hest measured value 1 Select the desired measurand 2 To display the maximum value of the selected measurand press the button for at least five seconds 3 1 To reset the instrument to its normal operati...

Page 10: ...or high humidity temperature calibration point REL1 REL2 Status Relay ERROR 01 04 see Hardware chapter 8 3 Self diagnosis and error messages 5 MEASURED VALUES from up to units EE31 A F EE31 B EE31 D...

Page 11: ...Software chapter 5 4 Calibration i i 1 Insert the sensor probe into the reference humidity 1 low calibration point and stabilise for at least 20 minutes 2 BUTTON S2 Pressing the button for 5 seconds s...

Page 12: ...as hing 10 BUTTON S1 store Pressing the button for 5 seconds stores the calibration value and the procedure is ended LED D2 flashes to indicate exiting of the calibration mode and the symbol CALIB HIG...

Page 13: ...ation point 6 Insert the sensor probe into the reference temperature 2 high calibration point and stabilise for at least 10 minutes 7 BUTTON S1 Pressing the button for 5 seconds starts the procedure f...

Page 14: ...re The calibration mode is indicated by the lit LED D2 and the symbol CALIB HIGH will appear on the optional LCD display CP 50 RH or BUTTON S2 Pressing the button for 5 seconds starts the procedure Th...

Page 15: ...ressing the button for 5 seconds starts the procedure The calibration mode is indicated by the symbol CALIB HIGH on the optional LCD display CP 45 degC or BUTTON S2 Pressing the button for 5 seconds s...

Page 16: ...seconds starts the procedure for the calibration mode RH Pressing both buttons simultanously for 10 seconds customer calibration settings are reset to factory calibration A short flash of the LED D1 i...

Page 17: ...Start configuration software on PC 9 For further instructions see Configuration software chapter 5 3 Sensor Probe replacement i active side Transmitters of the EE29 31 xDx and xEx series are availabl...

Page 18: ...transmitter Reset to factory calibration and repeat the calibration routine Filter soiled Replace filter Sensor defective Replace sensor Output configured incorrectly PC Software Long response time Fi...

Page 19: ...les external diameter 4mm 2 core twisted pair Typ 50 pF m impedance 100 Ohm non shielded In accordance with the RS485 standard cables in category 5 UTP specified according to EIA TIA ANSI 568 meet the...

Page 20: ...e following plug connectors are also necessary Y splitter Siemens 6ES7 194 1KA01 0XA0 Plug Lumberg RSC 5 7 RS232 485 Converter To adapt the RS232 interface on the PC to the RS485 network protocol a si...

Page 21: ...29 31 xEx with 10 m cable for EE29 31 xEx with 20 m cable Replacement sensors Replacement humidity sensor with sensor data Replacement humidity sensor with coating and sensor data Replacement temperat...

Page 22: ...rature Tf 80 112 0 32 0 32 0 32 degC F Wet bulb temperature Tw 0 32 60 140 80 176 100 212 degC F Water vapour partial pressure e 0 0 200 3 500 7 5 1100 15 mbar psi Mixture ratio r 0 0 425 2900 999 999...

Page 23: ...pectively 0 10V 1mA IL 1mA 4 20mA RL 500 Ohm 0 20mA RL 500 Ohm General Supply voltage SELV 8 48V DC SELV 12 35V AC SELV Self Electrical Low Voltage Current consumption 2x Voltage output for 24V DC AC...

Page 24: ...rement accuracy cannot be guaranteed Relative humidity RH Temperature degC Display graphical LCD 128x32 pixels with integrated push buttons for selecting parameters and MIN MAX function Alarm outputs...

Page 25: ...uirements MS WINDOWS 98 or higher RS232 serial interface 2 INSTALLATION Insert the CD ROM supplied with the transmitter into your PC and open the set up application Follow the instructions of the dial...

Page 26: ...tree structure only not to the configurations of individual transmitters Therefore only the network structure is saved Transmitters configurations must be reloaded using the command Read All Transmitt...

Page 27: ...ure requires the input of a number of parameters Group Assigns a transmitter to a group Network The check box network must be selected when several transmitters are operated in a network EE31 series I...

Page 28: ...will be displayed Other values are shaded and can not be selected or changed Write Writes the current configuration into the selected transmitter Write All Writes the current configuration to all sele...

Page 29: ...chapter 3 4 Transmitter Delete Transmitter see Configuration Software chapter 3 4 Transmitter 5 INDEX INDEX CARDS 5 1 Analogue For easy configuration of both analogue outputs Range Using the drop dow...

Page 30: ...High Sets the high switching point Hysteresis Sets the switching hysteresis that should be maintained each time the signal falls below the upper switching threshold Schaltpunkt switching point C kj k...

Page 31: ...ter 8 1 Sensor Replacement 3 Enter the nominal capacity C76 and the humidity coefficient in the corresponding input fields 4 Save the settings by clicking on the button Save Transmitter Replacement Pr...

Page 32: ...ate measurement results at a defined working point humidity point For calibration procedure see Hardware chapter 7 Humidity Temperature Calibration 1 Stabilise the probe of the desired humidity for mi...

Page 33: ...value in the input field Humidity Reading with the reference humidity Value of the saline solution or display of HUMOR 20 8 By clicking on Save the humidity reading of the transmitter will be adjusted...

Page 34: ...rmation on the date of the last humidity calibration but only where this was performed using the configuration software Note Manual calibration processes performed directly on the circuit board are no...

Page 35: ...M port serial interface of the PC Notebook in the pull down menu Interface Enter the name for the transmitter in the Name field Complete the New Transmitter process by clicking on the button Add 6 2 H...

Page 36: ...ce Building No 8 Bei Chen Dong St Chao Yang District Beijing 100101 P R China Tel 86 10 84992361 86 10 84992362 Fax 86 10 84992363 info epluse cn E E FRANCE Les Cr tes Dor es 1 F 69210 Sourcieux les M...

Reviews: