User Manual EE360
Moisture in Oil Sensor | 25
5.7 Pluggable Probe (Option PC4)
EE360 sensors are optionally available with pluggable sensing probe, which is attached to the EE360 enclosure
by a push-pull plug. If the probe or the probe cable gets damaged it is possible to easily replace the probe without
humidity and temperature adjustment. The replacement probe (see 8 Spare Parts and Accessories) is supplied
with a set of 7 individual parameters.
PLEASE NOTE
When using two or more devices at the same time, do not mix up the sensors. The serial number of the
connected sensor can be read by scanning the barcode on the basic unit.
1.
2.
2D-barcode for probe serial number
Fig. 20 Pluggable probe
Probe replacement procedure
▪
Plug off the damaged probe
▪
Plug on the new probe
▪
Update the 7 parameters with PCS10 Product Configuration Software or via display and push buttons (see
chapter 10 Appendix, Fig. 38 Device settings).
PLEASE NOTE
When replacing the probe, the factory calibration loses its validity. A calibration to verify accuracy is
recommended but not mandatory.
6 Setup and Configuration
The EE360 is ready to use and does not require any configuration by the user. The factory setup of EE360
corresponds to the type number ordered. For ordering guide please see datasheet at
.
This chapter describes how to customize the device.