background image

MANUAL

MANUAL

Serie EE99-1

Serie EE99-1

FFE

EU

UC

CH

HT

TE

E  //  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

R

FFE

EU

UC

CH

HT

TE

E  //  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

R

M

ME

ES

SS

SU

UM

MFFO

OR

RM

ME

ER

R

M

ME

ES

SS

SU

UM

MFFO

OR

RM

ME

ER

R

H

HU

UM

MIID

DIIT

TY

Y  //  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

RE

E

H

HU

UM

MIID

DIIT

TY

Y  //  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

RE

E

T

TR

RA

AN

NS

SM

MIIT

TT

TE

ER

R

T

TR

RA

AN

NS

SM

MIIT

TT

TE

ER

R

T

TR

RA

AN

NS

SM

ME

ET

TT

TE

EU

UR

R  D

DE

E  M

ME

ES

SU

UR

RE

E

T

TR

RA

AN

NS

SM

ME

ET

TT

TE

EU

UR

R  D

DE

E  M

ME

ES

SU

UR

RE

E

D

D''H

HU

UM

MIID

DIIT

TE

E  //  D

DE

E  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

RE

E

D

D''H

HU

UM

MIID

DIIT

TE

E  //  D

DE

E  T

TE

EM

MP

PE

ER

RA

AT

TU

UR

RE

E

BA_EE99-1_05 // Technische Änderungen vorbehalten / technical data are subject to change / sous réserve de toutes modifications techniques // 192057 

Summary of Contents for EE99-1 Series

Page 1: ...TE EM MP PE ER RA AT TU UR RE E T TR RA AN NS SM MI IT TT TE ER R T TR RA AN NS SM MI IT TT TE ER R T TR RA AN NS SM ME ET TT TE EU UR R D DE E M ME ES SU UR RE E T TR RA AN NS SM ME ET TT TE EU UR R...

Page 2: ...erference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measure...

Page 3: ...DIAGRAM 10 4 DIMENSIONS 10 5 HUMIDITY CALIBRATION 11 5 1 2 point humidity calibration 11 5 2 1 point humidity calibration 12 5 3 Reset to factory calibration 12 6 MAINTENANCE 12 6 1 Sensor exchange 1...

Page 4: ...heitshinweise an Sicherheitshinweise sind unbedingt zu befolgen Bei Nichtbeachtung k nnen Verletzungen von Personen oder Sachsch den entstehen E E Elektronik bernimmt daf r keine Haftung Dieses Zeiche...

Page 5: ...orgungsspannung 10 35V DC 10 28V AC T Sensor passiv 3 Draht 1 TASTER S2 f r 1 Pkt Feuchtekalibration Feuchte 50 r F f r 2 Pkt Feuchtekalibration unterer Kalibrationspunkt zum Verlassen der Kalibration...

Page 6: ...n wird der Messwert in 0 1 Schritten auf den Referenzwert abgeglichen Der Messwert kann entweder am Display abgelesen oder am Ausgang gemessen werden 4 TASTER S1 Durch mind 3 sek langes Dr cken wird d...

Page 7: ...S2 Durch mind 3 sek langes Dr cken wird die Routine verlassen ohne die Kalibrationswerte zu speichern Das Verlassen des Kalibrationsmodes wird durch das Deaktivieren der LED Calib und dem Symbol CAL i...

Page 8: ...F Genauigkeit inkl Hysterese und Nichtlinearit t mit Spezialkalibrierung gegen zertifizierte Standards 1 0 90 rF 2 90 100 rF Standardkalibration 2 0 90 rF 3 90 100 rF Ausgangssignal 4 20mA 3 Leiter An...

Page 9: ...ions should always be followed carefully By not following these instructions injuries of persons or material damage could happen Therefore E E Elektronik does not accept liability This symbol indicate...

Page 10: ...e 3 wire 1 PUSHBUTTON S2 for 1 point humidity calibration humidity 50 RH for 2 point humidity calibration low calibration point for exit the calibration mode 2 CALIB LED lit up permanently during the...

Page 11: ...1 up or down to the reference value The actual measuring value is indicated on the display or can be measured with the analogue output 4 PUSHBUTTON S1 Pressing the button for 3 seconds the calibration...

Page 12: ...ol CAL on the LC display or PUSHBUTTON S2 Pressing the button for 3 seconds the calibration procedure will be ended without storing the calibration values Exiting the calibration mode is indica ted by...

Page 13: ...tified standards 1 0 90 RH 2 90 100 RH standard calibration 2 0 90 RH 3 90 100 RH Traceable to intern standards administrated by NIST PTB BEV Output signal 4 20mA 3 wire Response time with filter at 2...

Page 14: ...s curit Les instructions de s curit doivent tre obligatoirement respect es Tout non respect peut provoquer la blessure des op rateurs ou l endommagement des mat riels E E Elektronik n assume alors au...

Page 15: ...E S2 touche pour la calibration 1 point d humidit humidit 50 HR touche pour la calibration 2 points d humidit point de calibration bas touche d abandon de la proc dure de calibration 2 DIODE LED reste...

Page 16: ...de 0 1 La valeur de mesure peut tre consult e sur l cran ou tre transmise via la sortie 4 TOUCHE S1 Apr s pression sur cette touche durant 3 secondes minimum la valeur de calibration est enregistr e...

Page 17: ...ression sur cette touche durant 3 secondes minimum a proc dure est abandonn e sans enregistrement des valeurs de calibration La sortie du mode de calibration est indiqu e par l extinction de la diode...

Page 18: ...justesse incluant hyst r sis et non lin arit Calibration sp ciale selon talons certifi s 1 0 90 HR 2 90 100 HR Calibration standard 2 0 90 HR 3 90 100 HR Tra abilit aux talons intern tels que NIST PTB...

Page 19: ......

Page 20: ...Hui Bin Office Building No 8 Bei Chen Dong St Chao Yang District Beijing 100101 P R China Tel 86 10 84992361 86 10 84992362 Fax 86 10 84992363 info epluse cn E E FRANCE Le Norly III 136 chemin du Mou...

Reviews: