background image

11

5.

HUMIDITY CALIBRATION

The EE99-1 transmitter series can be calibrated in two ways. 

-

1 point humidity calibration: quick and simple calibration on a defined humidity point 
(working point)

-

2 point humidity calibration: simple calibration for accurate measuring results over the
whole humidity working range.  

To reach a temperature balance it is recommended to keep the transmitter and the 
reference chamber (e.g. HUMOR 20,...) for minimum 4 hours in the same room. 

• 

During stabilisation period and calibration procedure it is important to keep the 
temperature constant in the reference climate chamber.

• 

For calibration the humidity sensor probe must be stabilised at least 30 minutes into 
the reference chamber.

• 

Replace a used dirty filter cap before calibration!

5.1

2 point humidity calibration

For accurate adjustment over the whole working range or in case of sensor exchanges a
two point calibration is recommended.

• 

Start calibration at the low humidity calibration point!

• 

The humidity difference between the two points should be > 30%RH

Procedure for 2 point humidity calibration 

(start at low calibration point)

:

1.

Insert the sensor probe into the reference chamber 1 (low humidity calibration point)

and stabilise for minimum 30 min.

2. PUSHBUTTON S2

: Pressing the button for 3 seconds starts the procedure for the

low calibration point. The calibration mode is indicated by the lit LED "Calib" and by the 
symbol "CAL" on the LC display.

3. PUSHBUTTON S1 (up) 

and

S2 (down)

: Pressing the two buttons will adjust the

measuring value in steps of 0.1% up or down to the reference value. The actual 
measuring value is indicated on the display or can be measured with the analogue output. 

4. PUSHBUTTON S1

: Pressing the button for 3 seconds the calibration value is stored

and the procedure is ended. Exiting the calibration mode is indicated by deactivation
of the LED "Calib" and the symbol "CAL" on the LC display.

or PUSHBUTTON S2

: Pressing the button for 3 seconds the calibration procedure will

be ended without storing the calibration values. Exiting the calibration mode is indica-
ted by deactivation of the LED "Calib" and the symbol "CAL" on the LC display.

5.

Insert the sensor probe into the reference chamber 2 (high humidity calibration point)

and stabilise for minimum 30 min.

6. PUSHBUTTON S1

: Pressing the button for 3 seconds starts the procedure for the

high calibration point. The calibration mode is indicated by the lit LED "Calib" and by
the symbol "CAL" on the LC display.

7. PUSHBUTTON S1 (up) 

and

S2 (down)

: Pressing the two buttons will adjust the

measuring value in steps of 0.1% up or down to the reference value. The actual 
measuring value is indicated on the display or can be measured with  the analogue output. 

8. PUSHBUTTON S1

: Pressing the button for 3 seconds stores the calibration value

and the procedure is ended. Exiting the calibration mode is indicated by deactivation
of the LED "Calib" and the symbol "CAL" on the LC display.

or PUSHBUTTON S2

: Pressing the button for 3 seconds the calibration procedure will

be ended without storing the calibration values. Exiting the calibration mode is indica-
ted by deactivation of the LED "Calib" and the symbol "CAL" on the LC display.

S1

S2

S1

S2

Calib

S1

S2

S1

S2

S2

S1

Calib

Calib

Calib

low calibration point:

high calibration point:

Summary of Contents for EE99-1 Series

Page 1: ...TE EM MP PE ER RA AT TU UR RE E T TR RA AN NS SM MI IT TT TE ER R T TR RA AN NS SM MI IT TT TE ER R T TR RA AN NS SM ME ET TT TE EU UR R D DE E M ME ES SU UR RE E T TR RA AN NS SM ME ET TT TE EU UR R...

Page 2: ...erference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measure...

Page 3: ...DIAGRAM 10 4 DIMENSIONS 10 5 HUMIDITY CALIBRATION 11 5 1 2 point humidity calibration 11 5 2 1 point humidity calibration 12 5 3 Reset to factory calibration 12 6 MAINTENANCE 12 6 1 Sensor exchange 1...

Page 4: ...heitshinweise an Sicherheitshinweise sind unbedingt zu befolgen Bei Nichtbeachtung k nnen Verletzungen von Personen oder Sachsch den entstehen E E Elektronik bernimmt daf r keine Haftung Dieses Zeiche...

Page 5: ...orgungsspannung 10 35V DC 10 28V AC T Sensor passiv 3 Draht 1 TASTER S2 f r 1 Pkt Feuchtekalibration Feuchte 50 r F f r 2 Pkt Feuchtekalibration unterer Kalibrationspunkt zum Verlassen der Kalibration...

Page 6: ...n wird der Messwert in 0 1 Schritten auf den Referenzwert abgeglichen Der Messwert kann entweder am Display abgelesen oder am Ausgang gemessen werden 4 TASTER S1 Durch mind 3 sek langes Dr cken wird d...

Page 7: ...S2 Durch mind 3 sek langes Dr cken wird die Routine verlassen ohne die Kalibrationswerte zu speichern Das Verlassen des Kalibrationsmodes wird durch das Deaktivieren der LED Calib und dem Symbol CAL i...

Page 8: ...F Genauigkeit inkl Hysterese und Nichtlinearit t mit Spezialkalibrierung gegen zertifizierte Standards 1 0 90 rF 2 90 100 rF Standardkalibration 2 0 90 rF 3 90 100 rF Ausgangssignal 4 20mA 3 Leiter An...

Page 9: ...ions should always be followed carefully By not following these instructions injuries of persons or material damage could happen Therefore E E Elektronik does not accept liability This symbol indicate...

Page 10: ...e 3 wire 1 PUSHBUTTON S2 for 1 point humidity calibration humidity 50 RH for 2 point humidity calibration low calibration point for exit the calibration mode 2 CALIB LED lit up permanently during the...

Page 11: ...1 up or down to the reference value The actual measuring value is indicated on the display or can be measured with the analogue output 4 PUSHBUTTON S1 Pressing the button for 3 seconds the calibration...

Page 12: ...ol CAL on the LC display or PUSHBUTTON S2 Pressing the button for 3 seconds the calibration procedure will be ended without storing the calibration values Exiting the calibration mode is indica ted by...

Page 13: ...tified standards 1 0 90 RH 2 90 100 RH standard calibration 2 0 90 RH 3 90 100 RH Traceable to intern standards administrated by NIST PTB BEV Output signal 4 20mA 3 wire Response time with filter at 2...

Page 14: ...s curit Les instructions de s curit doivent tre obligatoirement respect es Tout non respect peut provoquer la blessure des op rateurs ou l endommagement des mat riels E E Elektronik n assume alors au...

Page 15: ...E S2 touche pour la calibration 1 point d humidit humidit 50 HR touche pour la calibration 2 points d humidit point de calibration bas touche d abandon de la proc dure de calibration 2 DIODE LED reste...

Page 16: ...de 0 1 La valeur de mesure peut tre consult e sur l cran ou tre transmise via la sortie 4 TOUCHE S1 Apr s pression sur cette touche durant 3 secondes minimum la valeur de calibration est enregistr e...

Page 17: ...ression sur cette touche durant 3 secondes minimum a proc dure est abandonn e sans enregistrement des valeurs de calibration La sortie du mode de calibration est indiqu e par l extinction de la diode...

Page 18: ...justesse incluant hyst r sis et non lin arit Calibration sp ciale selon talons certifi s 1 0 90 HR 2 90 100 HR Calibration standard 2 0 90 HR 3 90 100 HR Tra abilit aux talons intern tels que NIST PTB...

Page 19: ......

Page 20: ...Hui Bin Office Building No 8 Bei Chen Dong St Chao Yang District Beijing 100101 P R China Tel 86 10 84992361 86 10 84992362 Fax 86 10 84992363 info epluse cn E E FRANCE Le Norly III 136 chemin du Mou...

Reviews: