User Manual HTS801
Humidity and Temperature Sensor for High Humidity and Chemically Polluted Conditions | 15
Each point in the graph represents a logged value. The points are connected by linear interpolation.
One point
Fig. 12 Data logger graph
The data logger memory works according to the first in first out (FIFO) principle: new data is stored while the
oldest.data.is.deleted..The.latest.20 000.logged.values.are.available.in.the.internal.memory.
The logged data can be downloaded with PCS10 Product Configuration Software as .csv file by choosing the
measurands and the time period.
Status Information
The status information shows all actual HTS801 settings and measured values.
Fig. 13 Status information
Buzzer
If a previously configured limit value is violated, a buzzer sounds periodically. This buzzer can be switched off or
on here. Factory default is on.
Error Indication
As soon as an error occurs, an indication shows the error code.
3.5.2.
rapidX Smart Probe (Option PC7)
The HTS801 with PC7 option features the intelligent pluggable rapidX probe. The rapidX smart probe enables
a plug-and-play probe exchange during operation without any configuration, adjustment or calibration. The
sensing element data of the rapidX smart probe is determined in the E+E RH factory calibration systems. Upon
connection to the HTS801, the probe automatically uploads its specific parameters to the HTS801 electronics.
This enables simple maintenance procedures at the customer’s site without “down time” due to cost-intensive
recalibration/repair.and.even.without.manual.parameterization.of.the.new.probe..The.smart.rapidX.probe.enables.
quick replacement without any tools at any time, even during operation (hot-swap).