User Manual HTS801
Humidity and Temperature Sensor for High Humidity and Chemically Polluted Conditions | 51
Ordering Guide HTS801 Probe (rapidX Smart Probe, PC7 option only)
Sensing element protection
Without, for very low temperature applications only
-70 °C...-40 °C (-94 °F...-40 °F)
C0
E+E proprietary coating
C1
HTS801P-
Type
Remote probe up to 180 °C (356 °F)
T5
Remote probe, pressure-tight up to 20 bar (300 psi)
and 180 °C (356 °F)
T10
RH remote probe, pressure-tight up to 20 bar (300 psi) and
180 °C (356 °F)
T17
Filter
Stainless steel sintered
F4
F4
PTFE (Polytetrafluoroethylene)
F5
F5
Stainless.steel.grid,.stainless.steel.body.(180.°C./.356.°F)
F9
F9
F9
PTFE membrane, stainless steel body
F11
Probe cable length
2 m (6.6 ft)
K2
K2
K2
5 m (16.4 ft)
K5
K5
K5
10 m (32.8 ft)
K10
K10
K10
Probe length
65 mm (2.56")
L65
L65
80 mm (3.15")
L80
200.mm.(7.87ʺ)
L200
L200
L200
400 mm (15.75")
L400
L400
L400
Process connection
G1/2".ISO.-.sliding.fitting,.Ø13.mm.(0.51")
PA23
1/2".NPT.-.sliding.fitting,.Ø13.mm.(0.51")
PA25
Feature
Description
Code
9 Technical Data
Measurands
Relative Humidity (RH)
Measuring range
0...100 %RH
Accuracy
1)
incl. hysteresis, non-linearity and repeatability
-15...+40 °C (5...104 °F) RH ≤90 %
-15...+40 °C (5...104 °F) RH >90 %
-25...+70 °C (-13...+158 °F)
-40...+180 °C (-40...+356 °F)
-70... -40°C (-94...-40 °F)
±(0.95 + 0.0013 * mv) %RH
±1.8 %RH
±(1.05 + 0.0084 * mv) %RH
±(1.15 + 0.013 * mv) %RH
±3.85 %RH
mv = measured value
Factory calibration uncertainty
2)
0...90 %RH
90...100 %RH
±(0.7 + 0.003 * mv) %RH
±1 %RH
mv = measured value
Temperature dependence of electronics, typ.
±0.01.%.RH./.°C.(0.0056.%RH./.°F)
Response time t
90
with metal grid filter at 20 °C (68 °F)
<15 s
1) Defined against E+E calibration reference.
2) Defined at 23 °C with an enhancement factor k=2, corresponding to a confidence level of 95 %.