Removing the
cartridge with filter (video available
on your personal customer page).
> Once the mobile head is removed, the
cartridge
with filter is then visible in the applicator (
2
).
> Remove the cartridge from the applicator: press fir-
mly on the filter support (
3
) and pull (
4
). The
cartridge with filter now presents itself to you (
5
); it is
imperative to hold it only by the plastic parts.
•
Cleaning of the internal optical parts.
> Once the filter cartridge is removed, the transparent
optical glass is also accessible (
6
).
> Using the spray
: spray it on a lint-free cloth
beforehand (
7
).
NOTE:
Never spray directly on the optical parts.
> Afterwards clean only the accessible side of the filter
with this cloth (
8
). Repeat the same procedure for the
accessible part of optical glass (
9
).
NOTE:
The filter must never be separated from its
cartridge.
> When cleaning is complete, take the
cartridge
and align it with the applicator with the help of the red
dots (
10
). Push it into the slide way of the applicator by
pressing on it (
11
), so as to ensure that it is properly in
position (
12
).
> Re-attach the mobile part of the applicator by pres-
sing simultaneously on the locking tabs of the mobile
head (
13
), making sure to align the internal red dot of the
head with the red dot of the applicator.
NOTE:
For better efficiency of the
, t is recommended to clean the filter before each use with the spray
provided when you purchased the
. Any damaged filter due to non-compliance with the
procedure for use is not covered by the manufacturers warranty.