background image

3.

 

Air Duster, aerosol can of compressed air. The preferred air duster is designed to leave no residue with a high 
filtration. The non-flammable model of the product is preferred. The following Brand has been used and tested 
by E-Seek. Chemtronics Ultrajet 70. 

 

https://www.chemtronics.com/p-844-ultrajet-70.aspx

 

 
 
 

There are three parts to maintain the M500:

 

 

Verification

 

(Step 1-2)

 

 

Cleaning

 

(Step 3-5)

 

 

Calibration

 

(Step 6-7)

 

 
 
 
 
 
 
 
 
 
 
 
 
 

Step 1: 

Insert the Calibration Card with the arrow facing outward (away from the

M500

)

.

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Calibration card inserted with arrow 

facing outward 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for M500

Page 1: ...is procedure are detailed below with pictures included The particularly tricky part at least the first time someone does it is getting the air divider in place Our best piece of advice is to look closely at the photos in the manual and realize that the tool will almost snap into place when inserted properly The calibration card and air divider are replaceable however you should take care to keep t...

Page 2: ...Dust lines yellowing at top of image ...

Page 3: ...will be the previous calibration settings The new settings will not be displayed until the next scan These settings are used for color balance calculations within theM500 during normal scanning operation The secondary purpose of the White Calibration Card feedback in the determination of whether or not dust or debris may have accumulated on the M500 s mirrors The Verification Process occurs when t...

Page 4: ...f card scans The general guideline is for cleaning once a month for high usage and once every three months for low usage environments This process is very simple to perform as it is an automatic process once the cleaning card is inserted into theM500 Alternatively you can purchase a 3rd party cleaning card We have tested and recommend the KICTeam Waffletechnology Cleaning card available on Amazon ...

Page 5: ...n process should be performed in a periodic basis to assess if dust has degraded the image quality Every 10000 to 20000 cards scanned the mirror inside the M500 will need to be checked for dust and cleaned The Detection and Cleaning process for the M500 mirrors requires the following items 1 Calibration Card this card is designed and made by E Seek Item M500 CALIB 2 M500 path Air Diverter a plasti...

Page 6: ...rred The following Brand has been used and tested by E Seek Chemtronics Ultrajet 70 https www chemtronics com p 844 ultrajet 70 aspx There are three parts to maintain the M500 Verification Step 1 2 Cleaning Step 3 5 Calibration Step 6 7 Step 1 Insert the Calibration Card with the arrow facing outward away from theM500 Calibration card inserted with arrow facing outward ...

Page 7: ...lication Note the horizontal dust lines at top of the card If there is dust on the mirrors you will see a dust line as illustrated by the number 1 on the figure below on the left section number2 of the image The Left section number 2 is the uncorrected side of the image and the Right Section ...

Page 8: ...opening 2 Position the Air Divider under the top spring loaded line Grey Plastic 3 Align the Air Divider to the M500 side wall 4 Push the Air Divider completely into the M500 Scan path Step 4 Use of the Air Duster Insert the pressurized Air Duster nozzle into the M500 scan opening Air Duster operation Follow the operational directions on the Air Duster for use Typically you want to press the Air R...

Page 9: ...side image Step 6 Calibration At this point you need to calibrate the M500 without any dust present on the mirrors This is a simple and mostly automatic process Simply insert the Calibration Card into theM500 with the arrow FACING INWARD The M500 will accept the card into the unit and goes through the calibration process automatically Once the calibration is complete the Calibration Card will be e...