1264C User Manual
4. Observe that the front panel LED indicators all turn Green.
Note that the FAN LED indicator will illuminate RED while
the fan ramps up to full speed. After approximately 20
seconds the FAN LED should turn Green.
WARNING:
Option 71 (P/N 407676) is required for applications powered
by 230 VAC power. The 1264C is configured for 230 VAC
operation ONLY when Option 71 is ordered.
Monitoring
Basics
Refer to
Figures 3-1 and 3-2
while using
Table 3-1
to interpret
system monitor information. Refer to Section 7 Troubleshooting if a
system fault is detected.
Table 3-1, Monitoring Basics
WHAT YOU SEE
ON THE FRONT PANEL
WHAT IT MEANS
Green LED illuminated for:
+5V
+12V
+24V
-2V
-5.2V
-12V
-24V
VXI Voltages are within acceptable limits
+4.56V
≤
+5V Rail
≤
+5.50V
+10.95V
≤
+12V Rail
≤
+13.20V
+21.90V
≤
+24V Rail
≤
+26.40V
-2.20V
≤
-2V Rail
≤
-1.83V
-5.72V
≤
-5.2V Rail
≤
-4.75V
-13.20V
≤
-12V Rail
≤
-10.95V
-26.40V
≤
-24V Rail
≤
-21.90V
Green LED illuminated for:
FAN
Cooling fan is operating
Green LED illuminated for:
TEMP
Power supply intake air temperature
≤
60
°
C
EADS North America Defense
Operating The 1264C 3-3
Test and Services, Inc. © 1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com